page 12 of 24
Operating Manual RS 1107, RS 1113
rspro.com
Factory settings
IN,T
NO
Use current configuration
YES
Reset product to factory settings.
IN,T DONE
appears
in the display
4.2
Open the adjustment menu
1.
The product is switched off
2.
Press and hold the
Down button
.
3.
Press the
On/Off button
to switch on the product.
4.
Release the
On/Off button
after 1 second and then the
Down button
in order to
call up the
adjustment menu
. The display shows the first parameter.
Parameter
Values
Meaning
_
Zero point correction: offset
PR
.
OF
0.00
No offset
-5.00 .. 5.00
Offset active (in hPa
– at RS 1107)
-50.0 .. 50.0
Offset active (in hPa
– at RS 1113)
Gradient correction: slope
PR
.
SL
0.00
No slope
-5.00 .. 5.00
Slope active (in %)
Formula:
Zero point correction:
Displayed value = measured value
– offset
Gradient correction:
Display = (measured value - offset) * (1 + slope / 100)