H88.0.2X.DK2RS-1.0
Operating Manual
EN
RS 1107-UT
Stock number: 205-0963
RS 1113-UT
Stock number: 205-0964
rspro.com
Page 1: ...H88 0 2X DK2RS 1 0 Operating Manual EN RS 1107 UT Stock number 205 0963 RS 1113 UT Stock number 205 0964 rspro com ...
Page 2: ...glance 7 3 1 The RS 1100 manometer series 7 3 2 Display elements 7 3 3 Operating elements 8 3 4 Connections 9 4 Operation 10 4 1 Opening the configuration menu 10 4 2 Open the adjustment menu 12 5 Measurement Basics 13 5 1 Special functions 13 5 1 1 NVLL Tare function 14 5 1 2 F NE High resolution measurement with 0 1 Pa RS 1107 14 5 1 3 F NE High resolution measurement with 1 Pa RS 1113 15 5 1 4 ...
Page 3: ...ting Manual RS 1107 RS 1113 page 3 of 24 6 2 1 Battery indicator 17 6 2 2 Changing battery 17 7 Error and system messages 18 8 Technical data 20 8 1 RS 1107 20 8 2 RS 1113 22 9 Service 24 9 1 Manufacturer 24 ...
Page 4: ...ument is entrusted to the recipient for personal use only Any impermissible transfer duplication translation into other languages or excerpts from this operating manual are prohibited The manufacturer assumes no liability for print errors 1 3 Further information _ Software version of the product V1 1 or later _ For the exact product name refer to the type plate on the rear side of the product ...
Page 5: ... product can only be guaranteed if applicable safety precautions and the device specific safety instructions for this docu ment are observed If these notices are disregarded personal injury or death as well as property damage can occur DANGER Incorrect area of application In order to prevent erratic behaviour of the product personal injury and property damage the product must be used exclusively a...
Page 6: ...1113 Differential pressure 200 hPa 2000 hPa Max resolution 0 1 Pa 1 Pa Usual applications include precise measurements of filter condition gas flow pressure draught leak integrity dynamic pressure flow speed The pressure connection is made at the supplied interchangeable pressure connection ports with suitable hoses The product must only be used under the conditions and for the purposes for which ...
Page 7: ...elements _ Display _ Battery indicator Evaluation of the battery status _ Unit display Display of the units or Min Max Hold information text _ Main display Measurement of the current pressure or value for min max hold _ Auxiliary display Measurement of the current pressure in Min Max Hold mode _ Bar graph Trend display with the special function F NE ...
Page 8: ...on the product Activate deactivate lighting Long press Switch off the product Reject changes in a menu _ Up Down button Press briefly Display of the min max value Change value of the selected parameter Long press Reset the min max value of the current measure ment Both simultaneously Rotate display overhead display _ ...
Page 9: ...Long press 2s Start menu configuration ONF appears in the display Close menu changes are saved Long press 4s Depending on the selected special function Activa tion of the Tare function NVLL high resolution measurement F NE or rapid measurement with mean value AVR 3 4 Connections Universal connection Interchangeable pressure connection via G1 8 thread ...
Page 10: ...pears in the display Release the Function key _ Parameter Values Meaning _ Display unit UN T Pa only at RS 1107 HPa mBaR BaR only at RS 1113 PSI mmKG _ Activatable special functions NVLL Tare function available F NE High resolution measurement with 0 1 Pa RS 1107 or 1 Pa RS 1113 activatable AVR 0 02 AVR 0 05 AVR 0 10 Rapid measurement with mean value over 2 s 5 s 10 s activatable _ ...
Page 11: ...justed to the lowest value e g 20 00 20 00 hPa RS 1107 200 0 200 0 hPa RS 1113 _ Shut off time POFF OFF No automatic shut off 0 15 0 30 1 00 4 00 12 00 Automatic shut off after a selected time in hours and minutes during which no buttons have been pressed _ Backlight L TE OFF Backlight deactivated 0 15 0 30 1 00 4 00 Automatic shut off of the backlight after a selected time in minutes and seconds ...
Page 12: ...product 4 Release the On Off button after 1 second and then the Down button in order to call up the adjustment menu The display shows the first parameter Parameter Values Meaning _ Zero point correction offset PR OF 0 00 No offset 5 00 5 00 Offset active in hPa at RS 1107 50 0 50 0 Offset active in hPa at RS 1113 Gradient correction slope PR SL 0 00 No slope 5 00 5 00 Slope active in Formula Zero ...
Page 13: ...rpose 6x1 mm PVC Up to 5 bar rel vacuum suitable 6x1 mm PE Up to 10 bar rel vacuum suitable 6x1 mm PUR Up to 9 bar rel vacuum suitable 5 1 Special functions _ With the special functions that can be selected via the Configuration menu the device can be optimised for special measuring tasks After it is switched on the device starts up in standard measuring mode the relevant special function is start...
Page 14: ... adjusted pressure applications In the Configuration menu the special function FVNC F NE has been selected The high resolution measurement can be activated by pressing and holding the Func tion key for 4 s Then the sensor is immediately zeroed and the optimised parameters for this measurement are activated CAUTION When starting the special function make sure that there is no pressure at the connec...
Page 15: ... and the optimised parameters for this measurement are activated CAUTION When starting the special function make sure that there is no pressure at the connec tions NOTE The increased current consumption in this mode decreases battery life The quickly determined measurement replaces other devices such as a U tube ma nometer The four bars in the lower display provide additional support The two middl...
Page 16: ...himney draft tests of forced air burners and consequently conventional electronic manometers are not adequate for task This special function optimises the device for this application purpose The different mean value times of 2 5 or 10 seconds can be selected depending on the requirement _ The first parameter is shown in the auxiliary display _ By pressing and holding the Function key for 2 s the s...
Page 17: ...rs in the main display the battery voltage is no longer adequate for operation of the product The battery is fully depleted 6 2 2 Changing battery _ DANGER Danger of explosion Using damaged or unsuitable batteries can generate heat which can cause the batter ies to crack and possibly explode Only use high quality and suitable alkaline batteries _ CAUTION Damage If the batteries have different char...
Page 18: ... screws A and remove the cover 2 Carefully replace the two Mignon AA batteries B En sure that the polarity is correct It must be possible to in sert the batteries in the correct position without using force 3 The O ring C must be undamaged clean and posi tioned at the intended depth In order to facilitate assem bly and avoid damage a suitable grease can be applied 4 Fit the cover on evenly The O r...
Page 19: ...urement too high Product is defective Stay within allowable meas urement range Send in for repair ERR 2 Measuring range is undercut Measurement too low Product is defective Stay within allowable meas urement range Send in for repair ERR 3 Display range has been exceeded Incorrect display unit Incorrect resolution F NE Function active Correct setting Deactivate function ERR 4 Display range has been...
Page 20: ... 1700 hPa Pressure connection 2 hose connections interchangeable with G1 8 universal ports Measuring cycle FAST approx 25 measurements per second SLO approx 2 5 measurements per second Display 3 line segment LCD additional symbols illuminat ed white duration adjustable Standard function Min Max Hold Auto power Off function if activated the product switches off automatically Activatable special fun...
Page 21: ...attery life approx 1 mA slow measurement SLO Operating time approx 3000 h Battery indicator 4 stage battery status indicator Note for low battery voltage BAT LO Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regula tions of the Member States 2014 30 EU EMC Directive 2011 65 EU RoHS Applied harmonised standards EN 61326 1 2013 ...
Page 22: ...ure connection 2 hose connections interchangeable with G1 8 universal ports Measuring cycle FAST approx 25 measurements per second SLO approx 2 5 measurements per second Display 3 line segment LCD additional symbols illuminat ed adjustable white permanent illumination Standard function Min Max Hold Auto power Off function if activated switches the product off automatically Activatable special func...
Page 23: ...t Battery life approx 1 mA slow measurement SLO Operating time approx 3000 h Battery indicator 4 stage battery status indicator Note for low charge level BAT LO Directives and standards The devices conform to the following Directives of the Council for the harmonisation of legal regula tions of the Member States 2014 30 EU EMC Directive 2011 65 EU RoHS Applied harmonised standards EN 61326 1 2013 ...
Page 24: ...ating Manual RS 1107 RS 1113 rspro com 9 Service 9 1 Manufacturer If you have any questions please do not hesitate to contact us RS Components Limited Birchington Road Corby Northamptonshire NN17 9RS WEE GF0002ZR Contact ...