BASIC OPERATION
21/11/2016 Version No. 002
67
Bandwidth
BW key sets the resolution bandwidth (RBW). The resolution bandwidth and the
sweep time are related. Please take into account how the sweep time is effected by
the resolution bandwidth.
Resolution Bandwidth Setting (RBW)
Description
The RBW (Resolution Bandwidth) defines the width of the IF
(intermediate frequency) filter that is used to separate signal
peaks from one another. The narrower the RBW, the greater
the capability to separate signals at close frequencies. But it
also makes the sweep time longer under specific frequency
spans (the display is updated less frequently).
Operation
1. Press
BW
>
RBW[F1]
and set the RBW to Auto or Man.
2. Set the resolution bandwidth and unit for Man mode.
Mode:
Frequency Range:
Auto, Man
1MHz, 300kHz, 100kHz
Note
The manual RBW is only available when the
span
≤
10MHz. If the span is greater than 10MHz then the
RBW is automatically set to Auto.