2.3 - Used reagents
2.3.1 - Starrsed Rinse solution
Starrsed Rinse solution is a dedicated IVD product, developed exclusively for automatic rinsing of the
Westergren pipettes in Starrsed ESR analyzers.
This reagent is supplied in a 20- liter container (QRR 010934).
2.3.2 - Starrsed Diluent
Starrsed Diluent is a dedicated IVD product, developed exclusively for the automatic dilution of blood
samples in Starrsed ESR analyzers.
▪
Each blood sample is automatically diluted with Starrsed Diluent
▪
Starrsed Diluent is used after each sample for cleaning the needle and fill nozzle assembly
This reagent is delivered in a 5-liter container (QRR 010931).
The solution should be discarded if it becomes turbid.
If the Diluent does become turbid, replace the Diluent container.
2.3.3 - Deionized water
After each aspiration, the fill nozzle is flushed with de-ionized water. The water container is a 5-liter plastic
container and must be filled up when empty.
If the deionized water comes up with an alarm with a filled container add then one or two drops of Starrsed
Diluent to the deionized water to avoid reagent alarm.
2.3.4 - Starrsed Disinfectant
Starrsed Disinfectant is a dedicated IVD product, developed exclusively for automatic disinfection of the
waste system of Starrsed ESR analyzers.
This reagent is delivered in a 5-liter container (QRR 010947).
2.3.5 - Starrsed Cleaning Agent
Starrsed Cleaning Agent is used to remove protein deposition from the Westergren pipettes of Starrsed ESR
analyzers.
Starrsed Cleaning Agent needs to be prepared for a cleaning procedure which is used in level 4
maintenance.
Starrsed Cleaning agent is supplied in 100 ml bottles.
QRR010905: 1 bag of 2 bottles
Instrument description
18
MRN-164_2-EN Version 1.08 Starrsed TL
Summary of Contents for STARRSED TL