If the QC sample is linked with a Lab ID, the sample will be requested at LIMS with the Lab ID.
If the QC sample is used with the original Starrsed Control sample ID, the sample will be requested at
LIMS with the Starrsed Control sample ID.
▪
OFF (=Default setting): All QC samples will be processed without requesting at LIMS.
Send QC result to LIMS
▪
ON: QC results are send to the LIMS as a standard ESR result.
If the QC sample is linked with a Lab ID, the result will be send to LIMS with the Lab ID.
If the QC sample is used with the original Starrsed Control sample ID, the result will be send to LIMS
with the Starrsed Control sample ID.
Note: When the MECHATRONICS-01 or MECHATRONICS-02 protocols are used, the "Sample code" (or
"Sample type") flag is set accordingly to mark the QC samples.
▪
OFF (=Default setting): Results will not be send to LIMS, data is only available on the Starrsed RS.
Automatically remove linked QC ID after result
▪
ON (=Default setting):The link between Starrsed Control Sample ID and Lab ID will be deleted after a
useable result has been reported for this particular lab ID. For each QC sample a new link must be
created. This link will stay active if there is no result or a general ESR error is generated. This setting is
useful when the lab issues a new and unique Lab ID for every QC sample.
When ready press
Save settings
before you continue.
6.7 - Service screen
Starrsed program
92
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...