6.5.1 - Prime / Clean
All maintenance functions for the fluid system are grouped under button
Prime / Clean
.
After each reagent change, the fluid system must be primed to fill the relevant tubes with reagent and
remove air. This is also part of the daily start-up. Use the applicable button to perform the automatic priming
cycle for this reagent:
Prime:
•
Prime
Pipettes
:
After each measurement, the pipettes are washed and dried automatically.
•
Prime
Needle/nozzle
:
After each aspiration, the outer needle, sample probe and fill nozzle are washed.
•
Prime
Diluter
:
The Diluter prime cycle is 5 strokes of the syringe.
•
Prime
Nozzle with water
:
After each aspiration, the fill nozzle is flushed with de-ionized water.
•
Prime
Waste system
:
During a pipette rinse cycle, a small amount of disinfectant is flushed around the bottom of the pipette
and into the waste system.
Starrsed program
MRN-071-2-EN version 7.02 Starrsed RS
67