Purpose:
Maintenance level 1
Safety:
Bio Hazard area
Instrument
: Starrsed RS
Revision:
002, March 2019
We recommend that this procedure is carried out by qualified service engineers.
The following items need to be replaced annually:
1. All tubing
ESRI079200 and additional tube set
.
2. Waste pump motor
ESRI090920.
3. Waste pump cassette
ESRI090921
.
4. Blue Vacuum filter disc. Part no
QWLV040003
.
5. Fill block washer. Part no
ESRI030906.
6. Waste container filter disc
QWLV040001.
(only applicable if internal waste container is used)
7. Teflon tip of syringe on the Diluter assembly (*Only applicable for Diluter syringe QWLV030009, Diluter
syringe QWLV030021 has to be replaced after 2 years).
The following items need to be replaced annually:
1. Outer needle and sample probe (if needed)
2. Pipette valve bodies (84 pieces)
QTST040001
.
Check:
1. Adjustment of fill nozzle and rinse nozzle.
2. Check measurement unit with Control pipettes.
Work instruction Starrsed RS
MRN-071-2-EN version 7.02 Starrsed RS
249
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...