If there is no liquid flow and no reagent alarm:
1. One of the de-ionized water lines may be blocked or kinked.
2. Add one or drops of Starrsed Saline or Diluent to get some conductivity in the De-ionized water.
15.6.2 - Disinfectant
To disinfect the ESR analyser unit waste line, select from
Maintenance -> Prime / Clean
-> [Prime Disinfectant]
, the
vacuum pump starts, and sufficient liquid must be seen flowing through the thin tube connected to the side
of the rinse nozzle
After each wash cycle, approximately 0.5 ml of disinfectant will be flushed through the wash station.
If no disinfectant flows;
1. Unscrew the cap of the disinfectant bottle or cubitainer to check.
2. One of the disinfectant lines may be blocked or kinked.
15.7 - Needle system
As soon as the barcode is accepted or the ID number keyed in manually, the sample will be processed.
15.7.1 - Needle not in top position
E4
E4 is the error code and the text in the status line
Sample probe not in top position ! (home). Piercing
needle returns not at the home position after sampling a tube.
•
Check the home sensor.
•
Sample probe motor is faulty.
•
Sample probe motor driver on needle board is faulty.
•
Sample probe is blocked
•
Fatal error, call distributor.
E14
E14 is the error code and the text in the status line
Outer needle motor position error ! (home). Outer
needle did not reach the home sensor within a certain time limit.
•
Check home (top) sensor.
•
Outer needle (tube) motor is faulty.
•
Outer needle motor driver on needle board is faulty.
•
Motor is blocked.
•
Outer needle is blocked.
•
Fatal error, call distributor.
Troubleshooting
152
MRN-071-2-EN version 7.02 Starrsed RS
Summary of Contents for STARRSED RS
Page 1: ...User Manual Starrsed RS MRN 071 2 EN version 7 02 ...
Page 2: ......
Page 16: ...Document history overview 16 MRN 071 2 EN version 7 02 Starrsed RS ...
Page 177: ...Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 177 ...
Page 201: ...Abbott Appendix for Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 201 ...
Page 219: ...Place Fill nozzle Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 219 ...
Page 251: ...Work instruction Starrsed RS MRN 071 2 EN version 7 02 Starrsed RS 251 ...