AC3006

Portable humidity and temperature instrument

General description

The HygroPalm21 is a highly accurate hand-held indicator that displays relative humidity, tem-
perature and the dew or frost point.

The HygroPalm22-A is a highly accurate hand-held indicator that displays relative humidity, 
temperature and a calculated parameter such as the dew or frost point.

Programming

The basic settings of the devices such as display, units (Metric / English), language and probe 
settings will be done in the factory. Some of the settings can be changed afterwards using the 
HW4 Software (optional). 

These short instructions are limited to a description of the main functions and installation of the 
device. The detailed instruction manual can be found on the internet: 

www.rotronic.com

Connections

Probe inputs (HP22-A)

Digital input for all HC2 ROTRONIC probes

Service connector

Service connector (Mini-USB) for a AC3006 cable

Functions Key

HOLD: (only HP22-A)

Short press, the measured value is frozen.

ON / OFF:

Press more than 1 second to turn on and off the device.

MENU

Activates the internal menu. Press this key again to go back 
to the previous menu or to exit the menu.

UP / DOWN 

When the menu is active, use to navigate the menu, make a 
selection or change a value.

ENTER 

When the menu is active, use to navigate the menu, make a selection or change a value.

Practical advice for measuring humidity 

The most common source of error when measuring relative humidity is a difference between the 
temperature of the probe and the temperature of the environment. At a humidity condition of 
50 %RH, a temperature difference of 1°C results in an error of 3 %RH on relative humidity. 
When using the HP21/22 hand-held indicator, it is good practice to monitor the display for tempe-
rature stability. The probe should be given sufficient time to equilibrate with the environment to be 
measured. The larger the initial temperature difference between the probe and the environment 
to be measured, the more time temperature equilibration requires. This time can be shortened, 
and errors avoided, by using the probe configuration that fits best for your application. In extreme 
situations, condensation may occur on the sensors when the probe is colder than the environment. 
As long as the humidity / temperature limits of the humidity sensor are not exceeded, condensation 
does not alter the calibration of the sensor. However, the sensor has to dry out before it can provide 
a valid measurement. Non-moving air is an excellent insulator. When there is no air movement, 
surprising differences in temperature and humidity can be noted over short distances. Air mo-
vement at the probe generally results in measurements that are both faster and more accurate. 

Changing the parameters shown on the display 

The temperature unit of the HP21 (°C or °F) HP22-A (metric or English)
can be changed from the keypad MENU > 

Device Settings > Unit. 

•  Relative humidity and temperature
•  Dew / frost point and temperature (HP21)
•  Relative humidity, temperature and calculated parameter (HP22-A)

Note HP22-A:

 As soon the calculated value is shown in the display, the parameter can be changed 

by using the UP / DOWN bottons  

Changing the units

The temperature unit of the HP21 (°C or °F) HP22-A (metric or English)
can be changed from the keypad MENU > 

Device Settings > Unit. 

•  Press the MENU key to show the internal menu on the display 
•  With the menu item Settings highlighted, press the ENTER key 
•  With the menu item Unit highlighted, press the ENTER key 
•  Use the UP or DOWN key to change the temperature unit as desired

Press the MENU key twice to exit the menu and return the HP21 to normal operation 

Low battery indicator

When the battery is down to about 20% of its initial charge, “Low Battery” appears at the bottom 
of the display.

Periodic calibration check of the probe

Both the PT100 RTD temperature sensor used in the probe and associated electronics are very 
stable and should not require any calibration after the initial factory adjustment. Long term stability 
of the ROTRONIC Hygromer

®

 humidity sensor is typically better than 1 %RH per year. For maximum 

accuracy, calibration of the probe should be verified every 6 to 12 months. Applications where the 
probe is exposed to significant pollution may require more frequent verifications. 

Menu structure  

(English language file) 

MAIN MENU

SELECTIONS / INFORMATION NOTES

Device Settings
Units

Metric / English

Back Light

Key Press / On / Off

Display backlight mode

Contrast (HP22-A)

LC display contrast adjustment

Trend (HP22-A)

On / Off

Trend indication on the display 

Device Information
Version

Firmware version

Serial Nbr

Serial number

Address

RS-485 address

Type

Device type

Name

Device name

User defined

Sensor Test (HP21)

Humidity sensor status

Off / Good / SQ-Tuned / Bad

SHORT INSTRUCTION MANUAL

Power supply 

The HP22-A uses a standard 9V alkaline battery. 

Um die To insert (replace) the battery, turn the latching 
button counter-clockwise and pull out the battery 
holder.

HygroPalm21

Integrated probe

HygroPalm22-A

Exchangeable probe

Measured Parameters

HP21

HP22-A

Humidity / Temperature

Calculated Parameters

Dew- / Frostpoint

Wet bulb temperature (Tw)  

Enthalpy (H) 

Vapor concentration (Dv) 

Specific humidity (Q)

Mixing ratio by weight (R) 

Vapor concentration at saturation (Dvs

Vapor partial pressure (E)  

Vapor saturation pressure (Ew

All HygroClip2-series probes can be connected to the HygroPalm22-A probe input. A calibration 
of the device, after changing the probe is not necessary. 

Connection to a PC or Laptop

Connect the HP21 and HP22-A via 
the service cable AC3006 (optional 
available) to a Laptop or PC. Data 
analysing, setting changes are done 
with HW4.

Display and display modes

The LC display has a backlight which can be set to be on all the time or 
whenever a key is pressed. The backlight can also be disabled.  Using 
the HP21/22 MENU >

 Device Settings > Back Light using the up / down 

buttons and confirm it with ENTER.

HP21:

 The first line of the display shows, depends on the settings, the humidity, dew- or frost 

point, on the second line the temperature.

HP22-A:

 The first line shows humidity, the second line the temperature and the third line the 

calculated value. 

Probe Information (HP22 only)
Version 

Firmware version

Serial Nbr 

Serial number

Address 

RS-485 address

Name 

Device name

User defined

SensorTest 

Humidity sensor status

Off / Good / SQ-Tuned / Bad

Record

On / Off

Data recording by the probe
(max. 2000 values)

Humidity Adjust
RefValue

Humidity reference value

±0.1 %RH steps

Acquired (HP22-A)

Number of cal. points in probe 
memory

<Acquire> (HP22-A)

Save cal. point to probe memory

<Delete> (HP22-A)

Erases all calibration points

<Adjust> (HP21)

Effect depends on number of 
calibration points

<Adjust> (HP22-A)

1-point adjustment only (offset)

Temperature Adjust
RefValue

Temperature reference value

±0.1 ˚C steps

<Adjust>

1-point adjustment only (offset)

Technical data

Battery type:  

9 V alkaline

Measuring range HP21

Humidity / accuracy: 

0…100 %RH / ±1.0 %RH @ 23 °C

Long term stability: 

<1 %RH/year

Temperature / accuracy: 

–10...60 °C / ±0.2 °C @ 23 °C

Long term stability: 

<0.1 °C/year

Maximum air velocity at probe: 

20 m/s

Measuring range HP22-A

Humidity and temperature: 

Depends on the probe

Environmental limits  

Storage and transit: 

–20…70 °C / 0…100 %RH, non condensing 

Operating limits at electronics: 

–10…60 °C / 0…100 %RH, non condensing

Dimensions: 

270 x 70 x 30 mm 

Weight: 

About 200 gr. 

The display can also be configured to show a trend indicator on each line: 

 increasing value

   ▼ 

decreasing value  

In the event of an alarm the symbol 

[ ! ]

 appears to the right of the value.


  

Constant value (End value is reached)

HygroPalm21/22-A

Summary of Contents for HygroPalm21

Page 1: ...der AUF AB Pfeile die Einheit C F beim HP21 und Metrisch Englisch beim HP22 A ausgewählt werden Durch drücken der ENTER Taste wird die Auswahl übernommen Durch zweimaliges drücken der MENU Taste gelangt man in den Normalmodus zurück Batterieladezustands Anzeige ErreichtdieBatterieeinenKapazitätwelcherkleinerals20 isterscheintimDisplay LowBatterie Periodische Kalibrierung des Fühlers Sowohl der Pt1...

Page 2: ...he display With the menu item Settings highlighted press the ENTER key With the menu item Unit highlighted press the ENTER key Use the UP or DOWN key to change the temperature unit as desired Press the MENU key twice to exit the menu and return the HP21 to normal operation Low battery indicator When the battery is down to about 20 of its initial charge Low Battery appears at the bottom of the disp...

Page 3: ... sur la touche ENTER Le retour en mode normal est effectué en appuyant deux fois sur la touche MENU Affichage de l état de la pile Le message Low Batterie apparaît si la capacité de la pile tombe à 20 de sa pleine charge Étalonnage périodique des capteurs transmetteurs de mesure L élément sensible de température Pt100 RTD ainsi que l électronique correspondante sont très stables et ne doivent norm...

Page 4: ...unità di misura C F nel HP21 e metrica inglese nel HP22 A utilizzando i tasti SU GIÙ Premendo il tasto ENTER si accetta la selezione Premendo due volte il tasto MENU si torna alla modalità standard Indicazione della carica della batteria Selabatteriaraggiungeunacapacitàinferioreal10 adisplayappareilmessaggio LowBatterie Calibrazione periodica della sonda Sia il sensore per la temperatura Pt100 sia...

Reviews: