background image

12

.1

00

1.

000

2

ROTRONIC AG

, CH-8303 Bassersdorf

Tel. +41 44 838 11 44, www.rotronic.com

ROTRONIC Messgeräte GmbH

, D-76275 Ettlingen

Tel. +49 7243 383 250, www.rotronic.de

ROTRONIC SARL

, 56, F - 77183 Croissy Beaubourg

Tél. +33 1 60 95 07 10, www.rotronic.fr

ROTRONIC Italia srl

I- 20157  Milano

Tel. +39 2 39 00 71 90, www.rotronic.it

ROTRONIC Instruments (UK) Ltd

, West Sussex RH10 9EE

Phone +44 1293 571000, www.rotronic.co.uk

ROTRONIC Instrument Corp

, NY 11788, USA

Phone +1 631 427-3898, www.rotronic-usa.com

ROTRONIC South East Asia Pte Ltd,

 Singapore 339156

Phone +65 6294 6065, www.rotronic.com.sg

ROTRONIC Shanghai Rep. Office

, Shanghai 200233, China

Phone +86 40 08162018, www.rotronic.cn

160

11

0

210

100

25

235

61

100

160

30

25

30

25

61

160

30

25

30

61

HF53S, HF54S (galvanically isolated), HF55S

Terminal Description

K2-1 V – 

Supply voltage – / Neutral

K2-2 V+ 

Supply v / Phase

K2-3  

Earth

K4-4 GND 

GND

K4-3 GND 

GND

K4-2 OUT2 

Analogue temperature  

*

K4-1 OUT1 

Analogue humidity  

**

K1-1 V+ 

DC (+) 17…24 VDC (+) See remarks below "K1"

K1-2 GND 

GND

K1-3 D + 

RS-485 Bi-directional TX + / RX +

K1-4 D – 

RS-485 Bi-directional TX – / RX –

*

  For humidity and calculated value output settings: 

  OUT2 = calculated value, OUT1 = humidity

**

  For temperature and calculated value output settings: 

  OUT1 = calculated value, OUT2 = temperature

Terminal K2-3:

 Earth is usually not connected to GND. If this is wanted, a land on the PCB must 

be removed. 

Terminals K1 (RS-485):

 Terminals K1-3 and K1-4 can be used to feed the device (multi-point 

connection). Several RS-485 devices can be operated with a strong 15 VDC power supply unit. In 
this case the supply voltage at K2-1 to K2-2 is not used.

Warning:

 Make sure that all settings have been made correctly before integrating and connecting 

the transmitters in the network.

Programming 

The basic settings of the devices are made in the factory according to your order. The transmit-
ters are adjusted in the factory and therefore do not need to be checked and readjusted during 
installation. The devices can be started immediately after installation.

Display 

In models with LC display the value can be read directly.

A

Digital transmitter for humidity & temperature
Duct & Wall Version

Congratulations on your purchase of the new state-of-the-art HygroFlex5-Series transmitter. Please 
read these short instructions carefully before installing the device

General description

The HygroFlex5-Series devices are universal transmitters for transmission of humidity and tempera-
ture measurements. Compatible with all interchangeable HC2 probes. These short instructions are 
limited to a description of the main functions and installation of the device. The detailed instruction 
manual can be found on the internet at: 

www.rotronic.com

Dimensions / Connections

Duct version

Wall version

Mechanical installation
General recommendations

Relative humidity is extremely temperature-dependent. In order to measure it exactly, the probe and 
sensors must be set exactly on the temperature level of the environment that is to be measured. 
The installation site can therefore have a significant influence on the performance of the device. 
Follow the guidelines below to ensure optimum performance:

a)  Select a representative installation site: Install the probe at a point where the humidity, tem-

perature and pressure conditions are representative for the environment that is to be measured. 

b)  Make sure there is sufficient air movement around the probe: An air flow of at least 1 metre/

second accelerates and facilitates adjustment of the probe to changing temperatures. 

c) Avoid:
  1. Probe too close to heating elements, cooling coils, cold or hot walls, direct sunlight, etc.
  2. Probe too close to steam, injectors, humidifiers or direct precipitation.
  3. Unstable pressure conditions with high air turbulence.
d  Insert the probe as far as possible into the environment that is to be measured.
e)  Avoid accumulation of condensation at the contact wires of the sensor. Install the probe so 

that the tip points down. If that is not possible, install it in horizontal position. 

Mounting the duct version 

To avoid measurement errors, at least 200 mm of the probe should be inserted into the environ-
ment that is to be measured. 
If necessary, use the mounting flange AC1305-M + AC1307 to install the probe and fasten the 
transmitter. 

AC1305-M + AC1307 Mounting flange
for 15...16 mm probe 

Mounting the wall version

Alignment

Mount the transmitter so that the probe points down.
 

Mounting variant 1

Drill the necessary holes using the drill template drawn on the packaging. Then insert the plugs 
delivered with the device and mount the transmitter with the screws. 

Electrical installation
Power supply

a)  HF53 (3-wire with analogue outputs): 6 to 40 VDC or 6 to 28 VAC.

 

Max. power consumption: <3.8 Watt

b)  HF54 (3-wire galvanic separated with analogue outputs): 9 to 36 VDC or 7 to 24 VAC.
  Max. power consumption: <3.8 Watt.
c)  HF55 (3-wire with digital output: 17 to 40 VDC or 14 to 28 VAC.
  Max. power consumption: <3.8 Watt

Supply voltage / Technology
Type 

Supply voltage V+ 

Load 

Output

3 / 4-wire

HF531S 

7...40 VDC / 4...  28 VAC 

Max. 500 

Ω

 

0...20 mA

HF532S 

7...40 VDC / 4...28 VAC 

Max. 500 

Ω

 

4...20 mA

HF533S 

6...40 VDC / 6...28 VAC 

Min. 1000 

Ω

 

0...1 V

HF534S 

10...40 VDC / 8...28 VAC 

Min. 1000 

Ω

 

0...5 V

HF535S 

15...40 VDC / 12...28 VAC 

Min. 1000 

Ω

 

0...10 V

Galvanically separated

HF541S 

9...36 VDC  

Max. 500 

Ω

  

0...20 mA

HF542S 

9...36 VDC  

Max. 500 

Ω

 

4...20 mA

HF543S 

9...36 VDC  

Min. 1000 

Ω

 

0...1 VDC

HF544S 

9...36 VDC  

Min. 1000 

Ω

 

0...5 VCD

HF545S 

9...36 VDC  

Min. 1000 

Ω

 

0...10 VCD

Digital output

HF556S 

17...40 VDC / 14...28 VAC 

 

Digital output

HF557S 

Power over Ethernet (PoE) 

 

Digital output 

 Caution: 

Wrong supply voltages and excessively high loading of the outputs can 

damage the transmitter.

Terminal configuration / Connection diagrams

The type is defined using the table Supply voltage / Technology to then use the following connec-
tion diagrams:

Button MENU: 

open / close menu  

Button ENTER: 
select menu point 

Menu navigation: 

B / - change value: increase/decrease

Note:

 Unauthorised use of the menu can be prevented by locking the setting “Display Menu” 

(using the HW4 software > Device Manager > Display).

The main menu points 
Main menu 

Menu points 

Options / Information 

Remarks

Decive Settings

 

Units  

Metric / English 

 

Contrast  

 

LC display contrast  

    

adjustment

 

Trend  

On / Off 

Trend shown on the  

    

display

Device Information

 

 

Version  

Firmware version

 

Serial No. 

Serial number 

 

Address  

Address RS-485 

 

Type  

Device type 

 

Name  

Device name 

User-defined

Sources of error

Measured values can be influenced by the following factors: 

Temperature errors :

Adaptation time too short, cold outside wall, heating elements, sunlight, etc.

Humidity errors:

Steam, water spray, dripping water or condensation at the sensor, etc. Repeatability and long term 
stability are, however, not influenced by these factors even if the probe is exposed to high humidity 
or saturation with steam (condensation) over a longer period of time.

Soiling:

By dust in the air. The choice of probe filter depends on the amount of soiling  at the measuring 
point. The filter must be cleaned or replaced periodically.

Scaling / Adjustment / Firmware update

The following settings can be made with the help of the HW4 software and either the service cable 
AC3006 or AC3009:
•  new scaling of the outputs
•  adjustment
•  firmware update 
You can find a detailed description in the manual that you can download from our web site at 

www.rotronic.com

Periodic calibration of the probe / transmitter

Both the Pt 100 RTD temperature sensor and the corresponding electronics are very stable and do 
not normally need to be changed or calibrated after factory calibration. The long term stability of the 
ROTRONIC Hygromer humidity probes is typically better than 1 %rh per year. For maximum accuracy 
we recommend calibration of the probe about every six to 12 months. More frequent calibration 
can be necessary in applications where the sensor is exposed to pollutants. The calibration can 
be performed by the user himself on site or in the laboratory / workshop. For routine calibrations 
the probe should be checked at one or two points. 

The electronics of the transmitter do not normally require calibration in the field. They can be 
checked easily with the help of the probe simulator in the HW4 software package. The electronics 
cannot be repaired in the field and should be returned to the manufacturer in the case of problems. 
For details on calibration, please see the full version of the instruction manual, which you can 
download from the internet. 

Technical data

 (measurement)

Humidity:  

0...100 %rh

Temperature:  

–100...200 °C

Accuracy:  

Probe-dependent: ±0,8 %rh, ±0,1 K @ 23°C (HC2-S)

Protection: IP65
Outputs:  

Current or voltage signals, digital output depending on order code, 

 

UART service interface

Technical data 

(

Electronics operating range

)

Temperature:  

–40...60 °C (–10...60 °C with Display)

Humidity:  

0...100 %rh, non-condensing

SHORT INSTRuCTION MANuAL

Current output

Voltage output

=

~

K2: V +

K2: V 

K2: GND

K4: OUT1
K4: OUT2

K2: V+

K2: V 

K4: GND

K4: OUT1
K4: OUT2

Equipment depending on model

Digit

al 

Analog

 

Pow

er

 

 

 

Sup

ply

USB 

Ethernet

PoE

K 1

1
2
3
4

B5

K 4

4
3
2
1

1    2    3

K 2

Reviews: