As first step to achieve this, the offset value must be set correctly.
This is usually done at 35 %rh.
The offset is now adjusted, the blue line goes through the 35 % point. As the line is not
going trough zero and maximum, the slope and linearity must be adjusted next: The
slope will be set at 80 %. This results in linearity between 30 and 100 %.
(0) (10)
35
80
Output
Ideal curve
Before adjustment
Relative humidity
Adjustment points
After adjustment
(0)
(10)
35
80
Pivot at 35%
Output
e.g. 80%
Relative humidity
Adjustment points
\\Kalibrierung\Calibration_procedures_2.doc