LuciPac

®

 A3 Aqua 

From Kikkoman 

 

For analysis of contamination of liquids via ATP/ADP/AMP measuring by Lumitester 
PD30/Smart, e.g. for contamination monitoring of waters and waste waters. 

 

1.  Introduction 

Using the LuciPac

®

 A3 Aqua sampling pen, liquids can be tested quickly and easily for organic 

contamination. The organic contamination of liquids, e.g. by bacteria, is indicated by the presence 
of ATP/ADP/AMP in these samples. An enzyme/substrate complex in the LuciPac

®

 A3 Aqua 

sampling tube reacts with ATP/ADP/AMP to produce bioluminescence whose intensity correlates 
directly with the amount of ATP/ADP/AMP present. The Lumitester PD30 and Lumitester Smart 
(1E37.1) then determine the bioluminescence as a direct measure of original contamination. The 
sampling comb ensures precise sampling of 150 

μl with a deviation of ≤2 %. Test results are, 

therefore, highly reproducible and comparable. 
The unit is easy to operate, even by non-professionals. Just immerse the sampling comb into the 
liquid. Then put the comb back into the tube and push it down into the reaction chamber. Shake 
the tube until the reagents are released and measure the bioluminescence in the Lumitester PD30 
or Lumitester Smart. 

 

2.  Shipment and storage 

20 LuciPac

®

 A3 Aqua each are packed in a resealable aluminium bag. Kits are to be stored at a 

low temperature (2 °C to 8 °C (35.6 °F to 46.4 °F)) for long term storage. The kit can be stored 
below 25 °C (77 °F) for up to 14 days or below 30 °C (86 °F) for up to 5 days before opening an 
aluminum bag without any adverse effect on the long term stability. Do not freeze the device. 
We recommend that you use all 20 sampling devices in a single bag at one time after opening an 
aluminum bag. If you have leftover sampling devices that you must store after opening a bag, be 
sure to store them at the recommended low temperature (2 °C to 8 °C (35.6 °F to 46.4 °F)) and 
use them within two weeks after opening. 
Expiry date: printed on the label of the aluminum bag. Store protected from light.

 

 

3.  Comments on application, detection limits 

The sampling stick (comb) of this device can collect up to 0.15 ml of water and ATP/ADP/AMP can 
be measured from 10

-11

 mol/l to 10

-6

 mol/l using LuciPac

®

 A3 Aqua combined with Lumitester PD30 

or Lumitester Smart (1E37.1). Therefore, the minimum measurable level of bacteria would 
correspond to about 10

4

 cells/ml calculated from said ATP content of 10

-18

 mol in a single 

bacterium. Please be aware that you cannot detect very low concentration of bacteria by this 
method.  
For detection of very low amounts of bacteria, membrane filtering methods should be applied. 
Please filter some bigger amount of liquid (typically 100 ml) through a membrane filter. 
Subsequently, probe this filter with LuciPac

®

 A3 (1ENH.1) and measure using the PD30 or 

Lumitester Smart (1E37.1).  
 

Instructions for use

 

Summary of Contents for LuciPac A3 Aqua

Page 1: ...tored at a low temperature 2 C to 8 C 35 6 F to 46 4 F for long term storage The kit can be stored below 25 C 77 F for up to 14 days or below 30 C 86 F for up to 5 days before opening an aluminum bag without any adverse effect on the long term stability Do not freeze the device We recommend that you use all 20 sampling devices in a single bag at one time after opening an aluminum bag If you have l...

Page 2: ...rase pyruvate orthophosphate dikinase PPDK and a pyruvate kinase PK The method produces a defined amount of luminescence proportional to the amount of adenosine triphosphate ATP adenosine diphosphate ADP and adenosine monophosphate AMP in a sample Firefly luciferase emits light in the presence of ATP and luciferin The AMP generated from this reaction is converted back to ATP using the PPDK while e...

Page 3: ...ted environment 2 C to 8 C 35 6 F to 46 4 F High temperatures may cause product performance to drop 5 Do not subject the device or any part of it to direct sunlight for long periods of time Strong light may cause product performance to drop 6 Do not touch any of the parts inside the device particularly not any part of the sampling stick comb itself with a finger or other object before use Touching...

Page 4: ... sampling stick to the main body casing and push it all the way into the main body 5 Push firmly onto LuciPac A3 Aqua casing and agitate 6 Allow the leftover luminescent reagent to thoroughly dissolve 7 Insert the LuciPac A3 Aqua into the Lumitester to measure the results Press ENTER 8 Remove the LuciPac A3 Aqua case from the measuring chamber 8 Safety recommendations Please observe the following ...

Page 5: ...606 0 Fax 49 0 721 5606 149 info carlroth com www carlroth com The company is a limited partnership with headquarters in Karlsruhe reg court Mannheim HRA 100055 Roth Chemie GmbH with headquarters in Karlsruhe reg court Mannheim HRB 100428 is the personally liable partner Managing Director André Houdelet Sales tax identification number DE 143621073 sse 07 2021 ...

Reviews: