Introduction

Thank you for purchasing a NA100 Compressor Nebulizer. 
With proper care and use, your nebulizer will provide you 
with many years of reliable treatments. This unit operates on 
standard AC power. Treatments are delivered quickly, safely and 
conveniently making this unit ideal for all ages. We encourage 
you to thoroughly read this guidebook to learn about the 
features of your nebulizer. Your compressor nebulizer should 

be used under the supervision of a licensed physician and/or 
a  respiratory  therapist.  Together  with  your  physician  and/or 

respiratory therapist, you can feel comfortable and confident 
knowing that you are obtaining the most effective inhalation 
treatments for your respiratory condition.

NOTE: 

Your nebulizer is intended for use in treatment of asthma, 

COPD  and  other  respiratory  ailments  in  which  an  aerosolized 
medication is required during therapy. Please consult with your 
physician and/or pharmacist to determine if your prescription 

medication is approved for use with this nebulizer. For type, 
dose, and regime of medication follow the instructions of your 
doctor or respiratory therapist. 

This  device  fulfils  the  provision  of  the  EC  directive  93/42/EEC 

(Medical Device Directive) and the European Standard EN 

13544-1:2007+A1:2009 Respiratory therapy equipment - Part 1: 

Nebulizing systems and their components. 

  Please read this manual carefully before use and be sure to 

keep this manual.

Cautions

Please  use  general  safety  precautions  when  operating  your 

nebulizer. This unit should be used only for its intended purpose 
as described in this guidebook and with medications only under 
the supervision and instruction of your physician. Do not use the 
device in anesthetic or ventilator breathing circuits.

Product cautions

READ THE FOLLOWING BEFORE USING

• To avoid electrical shock: keep unit away from water.
• Do not handle the unit of power cord with wet hands.
• Do not immerse the unit in liquid.
• Do not use while bathing.
•  Do not reach for a unit that has fallen into water immediately 

unplug the unit.

•  Do not use the unit if it has any damaged parts (including plug), 

if it has been submersed in water or dropped. Promptly send 

the unit for examination and repair.

•  The unit should not be used where flammable gas, oxygen or 

aerosol spray products are being used.

•  Keep the air vents open. Do not place the unit on a soft surface 

where the openings can be blocked. 

•  If the medication cup is empty, do not attempt to operate the 

unit.

•  If any abnormality occurs, discontinue to use until the unit has 

been examined and repaired.

• The unit should not be left unattended while plugged in.
• Do not tilt or shake the unit when in operation.
•  Disconnect the unit from the electrical outlet before cleaning, 

filling and after each use.

•  Do  not  use  attachments  unless  recommended  by  the 

manufacturer.

• Do not disassemble or attempt to repair the unit.

•  Do  not  use  the  device  in  anesthetic  or  ventilator  breathing 

circuits. 

Operating cautions

•  Close adult supervision is highly recommended when the unit is 

used by children or invalids.

• Keep your eyes away from the output of medication mist.
•  The maximum capacity of the medication cup is 5 ml and should 

not be overfilled.

• Do not use this unit while operating a vehicle.
•  If  any  discomfort  or  abnormality  occurs,  stop  using  the  unit 

immediately.

• Do not use the device if the air tube is bent.
•  Pentamidine  is  not  an  approved  medication  for  use  with  this 

device.

Storage cautions

•  Do  not  store  the  unit  in  direct  sunlight,  high  temperature  or 

humidity.

• Keep the unit out of reach of small children.
• Always keep the unit unplugged while not in use

Cleaning cautions

•  Check  air  filter,  nebulizer,  mouthpiece  and  any  other  optional 

component before each use. Dirty or worn parts should be 
replaced.

• Do not immerse the unit in water. It may damage the unit.
• Disconnect the unit from the electrical outlet before cleaning.
•  Clean  all  necessary  parts  after  each  use  as  instructed  in  this 

guidebook.

•  Always dispose of any remaining medication in the medication 

cup after each use. Use fresh medication each time you use the 
device.

•  Do not store the air tube with moisture or medication remaining 

in the air tube. This could result in infection as a result of bacteria.

MEDICAL DISCLAIMER: 

This manual and product are not meant to be a substitute for 
advice provided by your doctor or other medical professionals.
Don’t use the information contained herein or this product 
for diagnosing or treating a health problem or prescribing any 
medication. If you have or suspect that you have a medical 
problem, promptly consult your doctor.
This instrument is covered by a 1 year guarantee from the purchase 
date. The guarantee is valid only on presentation of the guarantee 
card completed by the dealer confirming purchase date or the 
receipt. Nebulizer Components are not included. Opening or 
altering the instrument invalidates the guarantee. The guarantee 
does not cover damage, accidents or non-compliance with the 

instruction manual. Please contact Rossmax Service or distributors.

Product specifications

Power

AC 230V/50Hz or AC 220V/60Hz or 

AC 110V/60Hz

Power Consumption

≤ 130W

Sound Level

≤ 55 dBA (1 meter away from 

NA100)

Compressor Pressure Range > 30 psi (207 kPa)

Operating Flow Range

 

4.0 lpm

Operating Temperature 

Range

10ºC to 40ºC (50ºF to 104ºF)

Operating Humidity Range 10 ~ 90% RH

Operating Atmospheric 

Pressure Range

700-1060 hPa 

Storage Temperature Range -20ºC to 60ºC (-4ºF to 140ºF)

Storage Humidity Range

10 ~ 90% RH

Dimension (L x W x H)

280mmx190mmx100mm 

(11.02x7.48x3.93 inches)

Weight

1750g (without accessories)

Medication Capacity

5ml(cc)

Particle Size (MMAD)

≤ 2.4μm

Dv50 (Spraytec)

≤ 4.4μm

Average Nebulization Rate

Fully open Valve ≥ 0.4ml/min

(0.9% Saline Solution)

Closed Valve ≥ 0.22ml/min 

(0.9% Saline Solution)

Standard Accessories

Nebulizer Kit, Air Tube, Mouth

-

piece, Filters, Adult and child 

masks, Service manual

*Subject to technical modification without prior notice.

* Performance may vary with drugs such as suspensions or high 

viscosity, See drug supplier’s data sheet for further details.

A

Product identification 

1. Nebulizer Kit 

2. Nozzle

3. Angled Mouthpiece

4. Air Filter

5. Child Mask

6. Adult Mask

7. Air Tube

8. Power Switch

9. Air Filter Slot

10. Air Output 11. Air Intake

12. Power Cord 

The nebulization rates can be adjusted by the user in a very easy 

way  without  exchanging  parts.  Higher  nebulization  rate/fully 

open is for higher viscosity medications and higher breathing 
capacity user while lower nebulization rate with closed valve 

will be more appropriate for kids / infants with lower breathing 

capacity.

C

Assembling your nebulizer kit

Follow the cleaning instructions in this guidebook under 
“Cleaning technique” prior to using your nebulizer for the first 
time or after it has been stored for an extended period of time.

REMEMBER: 

Always unplug the compressor and make sure the 

power-switch is turned to the “OFF” position before cleaning, 
assembling and before or after each use.

1. Place the compressor on a flat, stable surface within reach.

2.  Gently twist and pull straight up on the lid of the nebulizer to 

separate into two parts (medication cup and cover).

3.  Make sure that the nozzle is properly installed on the upper 

cover. The stem inside the medication cup inserts into the 
tube of the nozzle.

4.  Add the prescribed amount of medication to the medication 

cup.

5.  Reassemble the nebulizer by carefully twisting the medication 

cup and cover together. Make sure that the two parts fit 
securely.

   

WARNING: 

The symbol on this product means that it’s an 

electronic product and following the European directive 

2012/19/EU the electronic products have to be disposed 

on your local recycling centre for safe treatment.

B

Valve Adjustable Technology

The proprietary adjustable valve is able to deliver medications of 
different viscosity level according to every user’s conditions and 
needs at ease. Our VA technology allows users to adjust different 

levels  of  nebulization  rate  ranging  from  0.22  –  0.4  ml/min  at 

consistent particle size.

D

Operating your nebulizer

The nebulizer is operable at up to a 45° angle. If the angle is 
greater than 45°, no aerosol will be generated.
1.  Attach one end of the air tube connector to the air output.
2.  Carefully attach the opposite end of the air tubing connector 

to the stem at the base of the nebulizer kit.

3.  Attach the angled mouthpiece or mask to the top of the 

nebulizer.

4.  The capacity of the medication cup is 2~5 ml.

NOTE:

 A 30-minute interval is recommended after each use. The 

compressor will automatically shut off if it becomes overheated. 

If / when this happens, immediately:
1.  Press the power-switch to the “OFF” position.

2.  Unplug the power cord from the outlet.
3. Allow the motor to cool for 30 minutes.
Before restarting the unit, make sure that the air vents are not 
obstructed.