If the unit is used for over 10,000 times,
the calibration reminding message (CAL)
will appear every time the unit is switched
on. It is recommended that calibration be
completed every two years. Only the man-
ufacturer or an authorized service techni-
cian is qualified for the calibration service.
Please contact the manufacturer, Rossmax
International Ltd. or your local distributors. (The PC Link socket
for calibration purpose is inside the battery compartment.)
IN0GA1020000000XX
The default is to have the back light on once the pressure is in-
flated over 30 mmHg, and throughout the measurement pro-
cess. To switch the back light off at any time, press the “ ”
key.
1. Rossmax Mandaus II are exclusively designed for blood pres-
sure measurement at the upper arm or thigh on healthy skin.
2. Please do not inflate to more than 300 mmHg.
3. Do not take your blood pressure for more than 3 minutes.
4. Wait for at least 2 minutes before repeated measurements.
12. Safety Information
1. The unit contains high-precision assemblies. Therefore, avoid ex-
treme temperatures, humidity, and direct sunlight. Avoid drop-
ping or strongly shocking the main unit, and protect it from
dust.
2. Clean the blood pressure monitor body and the cuff carefully
with a slightly damp, soft cloth. Do not press. Do not wash the
cuff or use chemical cleaner on it. Never use thinner, alcohol or
petrol (gasoline) as cleaner.
13. Cautionary Notes
14. Specifications
15.
EMC guidance and manufacturer’s declaration
Guidance and manufacturer’s declaration-electromagnetic emissions
The Mandaus II is intended for use in the electromagnetic environment specified below.
The customer or the user of the Mandaus II should assure that it is used in such an environment.
Emission test
Compliance
Electromagnetic environment-guidance
RF emissions CISPR 11
Group 1
The Mandaus II uses RF energy only for its internal function. Therefore, its RF emissions
are very low and are not likely to cause any interference in nearby electronic equipment.
RF emissions CISPR 11
Class B
The Mandaus II is suitable for use in all establishments, including domestic establish-
ments and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC 61000-3-2
Not applicable
Voltage fluctuations/flicker
emissions IEC 61000-3-3
Not applicable
Guidance and manufacturer’s declaration-electromagnetic immunity
The Mandaus II is intended for use in the electromagnetic environment specified below.
The customer or the user of the Mandaus II should assure that it is used in such an environment.
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment-guidance
Electrostatic dis-
charge (ESD) IEC
61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile. If
floors are covered with synthetic material, the relative
humidity should be at least 30%
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
Not applicable
Not applicable
Mains power quality should be that of a typical com-
mercial or hospital environment.
Surge IEC
61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
Not applicable
Not applicable
Mains power quality should be that of a typical com-
mercial or hospital environment.
Voltage Dips, short
interruptions and
voltage variations
on power supply
input lines IEC
61000-4-11
<5% UT(>95% dip in UT) for 0,5
cycle
40% UT(60% dip in UT) for 5
cycles
70% UT(30% dip in UT) for 25
cycles
<5% UT(>95% dip in UT) for 5 s
Not applicable
Not applicable
Not applicable
Not applicable
Mains power quality should be that of a typical
commercial or hospital environment. If the user of
the Mandaus II requires continued operation during
power mains interruptions, it is recommended that
the Mandaus II be powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60 Hz) mag-
netic field IEC
61000-4-8
3 A/m
3 A/m
Power frequency magnetic fields should be at levels
characteristics of a typical location in a typical com-
mercial or hospital environment.
NOTE: UT is the a.c. mains voltage prior to application of the test level.
Guidance and manufacturer’s declaration-electromagnetic immunity
The Mandaus II is intended for use in the electromagnetic environment specified below.
The customer or the user of the Mandaus II should assure that is used in such and environment.
Immunity test IEC 60601 test level Compliance level
Electromagnetic environment-guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms
150 KHz to 80 MHz
3 V/m
80MHz to 2,5 GHz
Not applicable
3 V/m
Portable and mobile RF communications equipment should be used
no closer to any part of the Mandaus II, including cables, than the
recommended separation distance calculated from the equation
applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1,2 √P
d = 1,2 √P 80MHz to 800 MHz
d = 2,3 √P 800MHz to 2,5 GHz Where P is the maximum output power
rating of the transmitter in watts (W) according to the transmitter manu-
facturer and d is the recommended separation distance in metres (m).
Field strengths from fixed RF transmitters, as determined by an electro-
magnetic site survey, a should be less than the compliance level in each
frequency range.b
Interference may occur in the vicinity of equipment marked with the fol-
lowing symbol:
NOTE1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, ama-
teur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromag-
netic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in
the location in which the Mandaus II is used exceeds the applicable RF compliance level above, the Mandaus II should be observed
to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or
relocating the Mandaus II.
b. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Recommended separation distance between
portable and mobile RF communications equipment and the Mandaus II
The Mandaus II is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer
or the user of the Mandaus II can help prevent electromagnetic interference by maintaining a minimum distance between portable and
mobile RF communications equipment (transmitters) and the Mandaus II as recommended below, according to the maximum output
power of the communications equipment.
Rated maximum output
power of transmitter / W
Separation distance according to frequency of transmitter / m
150 kHz to 80 MHz / d = 1,2 √P
80 MHz to 800 MHz / d = 1,2 √P
800 MHz to 2,5 GHz / d = 2,3 √P
0,01
0,12
0,12
0,23
0,1
0,38
0,38
0,73
1
1,2
1,2
2,3
10
3,8
3,8
7,3
100
12
12
23
For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be es-
timated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range applies.
NOTE2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from
structures, objects and people.
Power Source
: DC 3V Two AAA Batteries
Measurement Range
: Pressure: 0-299 mmHg;
Pulse: 30-240 beats/min
Accuracy
: Pressure:
0°C~17°C, 34°C~46°C: ±6 mmHg
18°C~33°C: ±3 mmHg
Pulse: ±5% of reading
Pressure Sensor
: Semi-conductor
Inflation
: Manual Inflation
Deflation
: Manually operated deflation valves
Auto-shut-off
: 1 min. after last key operation
Back light auto-shut-off : 15 sec. after last key operation
Operation environment : 0°C~46°C (32°F~115°F);
85% RH max
Storage environment
: -34°C~65°C (-30°F~149°F);
85% RH max
Dimensions
: 78(L) x 63(W) x 48(H) mm
Weight
: 90g (Mandaus II)
267g [Mandaus II+Rossmax Gauge
Cuff (M)]
Limited users
: Adult users
: Type BF
Device and cuff are designed to pro-
vide special protection against electri-
cal shocks.
*Specifications are subject to change without notice.
1
2
4
3
5
6
1
2
4
3
5
6
9. Back light Control
Err
Heartbeat rate detection error. The cuff is deflated
too rapidly. Please keep the deflation rate among 2-3
mmHg per second.
EE
A pressure of over 15 mmHg remaining for more than 3
minutes is detected. Please press the “ ” key to switch
the unit off, or the unit will automatically switch off
after 60 seconds.
E3
Data Error. Please remove and reload the batteries. If
error keeps occurring, return the device to the local
distributor or service center.
01
The cuff is inflated to a pressure over 299 mmHg. Please
open the valve immediately to rapidly exhaust the air in
the cuff.
10. Error Messages
1
2
4
3
5
6
11. Maintenance
3. Leaky batteries can damage the unit. Remove the batteries when
the unit is not used for a long time.
4. The unit should not be operated by children so to avoid hazard-
ous situations.
5. If the unit is stored near freezing, allow it to acclimate at room
temperature before use.
6. Mandaus II is not field serviceable. You should not use any tool
to open the device nor should you attempt to adjust anything
inside the device. If you have any problems, please contact who
you purchased Mandaus II from or please contact Rossmax Inter-
national Ltd.
7. To stop operation at any time, open the air release valve, and the
air in the cuff will be rapidly exhausted.
8. Please note that this is a home healthcare product only and it is
not intended to serve as a substitute for the advice of a physician
or medical professional.
9. Do not use this device for diagnosis or treatment of any health
problem or disease. Measurement results are for reference only.
Consult a healthcare professional for interpretation of pressure
measurements. Contact your physician if you have or suspect any
medical problem. Do not change your medications without the
advice of your physician or healthcare professional.
10. Electromagnetic interference: The device contains sensitive elec-
tronic components. Avoid strong electrical or electromagnetic
fields in the direct vicinity of the device (e.g. mobile telephones,
microwave ovens). These may lead to temporary impairment of
measurement accuracy.
11. Dispose of device, batteries, components and accessories accord-
ing to local regulations.
12. This monitor may not meet its performance specification if stored
or used outside temperature and humidity ranges specified in
Specifications.
www.rossmaxhealth.com
Instruction
Manual
Model:
GA102
MANDAUS II
EN
Edition: V
-3-P
A01-3310
Issue Date: 2011/05/13