Rosslare PYR-3011 Installation Instructions Download Page 3

 

0706-00

 

5.2

 

APS – Auto Power Save Mode (JP1) 

To save power, the PYR-3011 goes into sleep mode after sending 

an alarm. The time the device will be latent is set by the radio 
mode jumper, JP1. 

 

Normal

 (always 2 minutes between alarms) – no jumper. 

 

Dynamic

 (2 minutes retriggered between alarms) – jumper 

on. 

When the dynamic mode is set, an alarm event will be sent only if 
there were 2 minutes of silence prior to the current alarm. This 

setting is useful for places with a high level of traffic, such as 

factories, shops, etc. 
 

5.3

 

Back Tamper (R11) 

The PYR-3011 has both front and back tamper detection. Front 
tamper detection is on by default and it is activated when the 

cover is opened cannot be changed. The back tamper detection 

is activated when the detector is pulled off the ceiling and is 

disabled by default, to enable the back tamper detection cut the 
back tamper wire - R11 - which is marked on the PCB as a scissor 

print.  

5.4

 

Test Mode Switch (S3) 

This pushbutton is used for walk test or radio test, as follows:  

Walk Test

 – press for 1 second only; PIR functionality will be 

evaluated during the calibration process, for a 1 minute period. 

(After self check at power up the walk test is automatic enable for 
2 minutes.) 

Radio Test

 – press for more than 3 seconds; the red indicator 

illuminates. 10 Alarms events are transmitted with 4 second 

intervals for a total transmission time of approximately 40 seconds.  
This test is to insure clear RF passes between the detector and the 
control panel. 
In all modes the LED will remain activate at all trouble 
announcements. 
Refer also to panel manual for RF testing details. 

 

Figure 2 PCB front view 

6.

 

Testing the Detector 

The PYR-3011 PIR has a built-in walk test function where the LED of 

the PIR is enabled. This test is used to check the detection of the 

PIR and the coverage pattern. 

To perform a walk test: 

1.  Ensure all of the settings in the PIR are adjusted as necessary for 

the location according to the installation instructions above.  

2.  Insert battery in the closed case as described above. The LED 

flashes on for 2 seconds and off for 2 seconds for a period of 1 

minute. Then the LED turns off. At this point a walk test can be 
performed. As default, perform a walk test of 2 minute at the 

end of the self test. If more than 2 minute goes past, refer to 

paragraph 3. 

3.  Place a magnet over the right side of the lens for 1 second 

only. 

4.  With the LED enabled the LED flashes every time the detector 

detects motion. There is a two second wait period before the 

next detection. 

5.  It is recommended that the installer test the detection by going 

over the protected area and seeing that the detection 

pattern is good. 

6.  After 1 minute, the LED goes off. If a new walk test is needed, 

go again to paragraph 3. 

7.

 

Signaling Table 

The LED on the front of the PYR-3011 is used to send several signals to the user. The following table describes the signals for different activities: 

Activity LED 

Signal 

Warm-Up 

The LED flashes on for 2 seconds and off for 2 seconds for a period of 1 minute. If the warm-up is 
successful, the LED stops flashing and the system is ready for detection. 

Detect Condition  

The LED flashes on for 2 seconds and then turns off. 

PIR Problem 

The LED flashes on for 1 second and then off for 1 second. A PIR check is conducted every 3 hours. 

Temperature Problem 

The LED flashes on in short bursts. 

Weak Battery 

LED OFF and ON alternately during Alarm, Tamper event 

8.

 

Testing the Detector 

The PYR-3011 PIR has a built-in walk test function where the LED of 
the PIR is enabled. This test is used to check the detection of the 

PIR and the coverage pattern. 

To perform a walk test: 

1.  Ensure all of the settings in the PIR are adjusted as necessary 

for the location according to the installation instructions 

above.  

2.  Insert battery in the closed case as described above.

 

The 

LED flashes on for 2 seconds and off for 2 seconds for a 

period of 1 minute. Then the LED turns off. At this point a walk 

test can be performed. As default, perform a walk test of 1 
minute at the end of the self test. If more than 1 minute pasts, 

refer to paragraph 3. 

3.  Push the guide light according to the required LED mode. 

(Refer to setting test Mode, above) 

4.  With the LED enabled the LED flashes every time the detector 

detects motion. There is a two second wait period before the 

next detection. 

5.  It is recommended that the installer test the detection by 

going over the protected area and seeing that the 

detection pattern is good. 

6.  After 1 minute, the LED goes off. If a new walk test is needed, 

go again to paragraph 3. 

9.

 

Low Battery Supervision 

Prior to each RF transmission, the battery voltage is sampled. If 
the voltage is low (as indicated by the red blinking light) for 3 

sequential samples, a special message is sent. Once the battery 

level returns to the minimum preset value, fault transmissions 

cease. 

Reviews: