4.3.10 Time-out settings
Master communication time-out
The master communication time-out specifies the elapsed time without
contact before communication with the master is considered to be lost.
This value is stored in database
register 188
. The default value is set to
20 s.
Note!
The database value is given in number of "ticks" 255. One tick is equal to
278
s, and 255 ticks is equal to 71 ms. Thus the value stored in
database register 188 must be multiplied with 71 ms in order to obtain
the time-out value in number of seconds.
RTG Communication time-out
The RTG communication time-out is stored in database
register 249
.
The default value is set to 60 seconds. This is the value which specifies
the time interval without contact before communication with the RTG is
considered to be lost.
Summary of Contents for TankRadar Rex DAU 2100
Page 2: ......
Page 3: ...www rosemount tg com Service manual Second edition Copyright May 2013 Rosemount Tank Radar AB ...
Page 8: ...Rosemount TankRadar Rex Contents TOC 4 Service Manual 308012En Ed 2 May 2013 ...
Page 116: ...Rosemount Tank Radar Rex Chapter 4 Service 4 72 Service Manual 308012En Ed 2 May 2013 ...
Page 124: ...Rosemount Tank Radar Rex Chapter 5 Calibration 5 8 Service Manual 308012En Ed 2 May 2013 ...
Page 132: ...Rosemount Tank Radar Rex Chapter 6 Troubleshooting 6 8 Service Manual 308012En Ed 2 May 2013 ...
Page 134: ...Rosemount Tank Radar Rex Chapter 7 Spare Parts 7 2 Service Manual 308012En Ed 2 May 2013 ...
Page 146: ...Rosemount Tank Radar Rex Chapter 9 Technical Data 9 10 Service Manual 308012En Ed 2 May 2013 ...
Page 152: ...Rosemount Tank Radar Rex Index 1 6 Service Manual 308012En Ed 2 May 2013 ...
Page 153: ......