Reference Manual
00809-0100-4663, Rev BA
January 2010
Rosemount 8732
E-6
BROOKS SENSORS
Connect coil drive and electrode cables as shown in Figure E-3.
Model 5000 Sensor to
Rosemount 8732
Transmitter
Figure E-3. Wiring Diagram for
Brooks Sensor Model 5000 and
Rosemount 8732
Table E-4. Brooks Model 5000
Sensor Wiring Connections
BROOKS MODEL
5000
ROSEMOUNT 8732
TRANSMITTER
Rosemount 8732
Brooks Sensors Model 5000
1
1
2
2
17
17
18
18
19
19
This is a pulsed DC magnetic flowmeter.
Do not connect AC power
to the sensor
or to terminals 1 and 2 of the
transmitter
, or replacement of the
electronics board will be necessary.