I
NSTALLATION
748221-L Rosemount Analytical December 2000
11
Model 7D Thermal Conductivity Analyzer
Note:
Route recorder cable through a separate conduit, not with power cable.
Output selection and cable connections for voltage-actuated and current-actuated
devices are explained in Sections 2.3.1 through 2.3.5.
2.3.1 S
TANDARD
(N
ON
-
LINEARIZED
) V
OLTAGE
O
UTPUT
1. On the Master Board (Figure 2-6):
a. Verify that TB2-1 is jumpered to TB2-2.
b. Set S1 for desired voltage: 5V, 1V, .1V, or .01V.
c. Connect the recorder cable to the TB4 terminals labeled VOLT OUT: TB4-1(+)
and TB4-2(-).
Note
Take the usual precautions to avoid AC pickup. DO NOT GROUND EITHER
LEAD.
2. Connect the cable to input terminals of the recorder; ensure that polarity is correct.
3. Ground shield on one end only.
2.3.2 L
INEARIZED
V
OLTAGE
O
UTPUT
(O
PTIONAL
)
1. Verify that Voltage-Output Linearizer Board (PN 633756) is properly inserted in
J102.
2. On the Master Board (Figure 2-6):
a. Verify that TB2-1 is jumpered to TB2-4, and TB2-2 is jumped to TB2-5.
b. Set S1 for desired voltage: 5V, 1V, .1 V, or .01 V.
c. Connect the recorder cable to the TB4 terminals labeled VOLT OUT: TB4-1 (+)
and TB4-2 (-).
Note
Take the usual precautions to avoid AC pickup. DO NOT GROUND EITHER
LEAD.
3. Connect the cable to the recorder input terminals; ensure that polarity is correct.
2.3.3 I
SOLATED
4
TO
20
M
A C
URRENT
O
UTPUT
(O
PTIONAL
)
1. Verify that the Isolated 4 to 20 mA Current Output Board (PN 652816) is properly
inserted in J103.
2. On the Master Board (Figure 2-6):