1.0

About this guide

This guide provides instructions for installing the Rosemount 5708 3D Solids 
Scanner in small vessel connections using the manhole mounting tool kit. It 
does not provide instructions for mounting the scanner head, electrical 
installation, or configuration.

Refer to the Rosemount 5708 

Quick Start Guide

 and Rosemount 5708 

Reference Manual

 for more instructions. The manual and this guide are also 

available electronically on 

Emerson.com/Rosemount

.

Authorized personnel

All operations described in this document must be carried out by authorized, trained personnel only.

For safety and warranty reasons, any internal work on the devices must be carried out by 
manufacturer-authorized personnel only.

Warnings about misuse

Inappropriate or incorrect use of the device may result in hazards and application-specific malfunctioning, 
such as vessels overfill or damage to system components through incorrect mounting or adjustments.

If the device is used in a manner not specified in this document, the protection provided by the device will 
be impaired.

General safety instructions

Consider local and national electrical codes and all common safety regulations and accident prevention 
rules during installation.

Substitution of components may impair intrinsic safety.

For preventing ignition of flammable or combustible atmospheres, read, understand and adhere to the 
manufacturer’s live maintenance procedures.

Contents

Install the antenna . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 4
Rotate the antenna toward vessel center   . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 9

Summary of Contents for 5708

Page 1: ...Quick Start Guide 00825 0200 4570 Rev AA June 2017 Rosemount 5708 3D Solids Scanner Manhole Mounting Tool Kit ...

Page 2: ...rk on the devices must be carried out by manufacturer authorized personnel only Warnings about misuse Inappropriate or incorrect use of the device may result in hazards and application specific malfunctioning such as vessels overfill or damage to system components through incorrect mounting or adjustments If the device is used in a manner not specified in this document the protection provided by t...

Page 3: ...23 0001 A Installation adapter B Set screw C 2 mm hex key Dimensions are in inches millimeters 1 2 Required equipment Standard tools e g screwdriver pliers 60 mm wrench Wrench to secure the mounting plate bolts 2 mm hex key Two ropes of appropriate type and length see page 6 A B C 1x 3x 1x 4 10 104 Ø 1 73 44 ...

Page 4: ...June 2017 4 Quick Start Guide 2 0 Install the antenna Step 1 Verify the O rings are in place Step 2 Remove the nut from the neck tube 60 mm ...

Page 5: ...Quick Start Guide 5 June 2017 Step 3 Mount the installation adapter on the antenna Step 4 Place the mounting plate and nut on the scanner s mounting hole 3x H2 mm Ø 2 05 in 52 mm ...

Page 6: ...end through the nut mounting plate and mounting hole to the manhole Step 6 Tie the rope to the installation adapter Note Consider securing the antenna by tying an additional rope to the adapter from the manhole opening to prevent the scanner from hitting the side walls Step 7 Carefully lower the antenna into the manhole ...

Page 7: ...Quick Start Guide 7 June 2017 Step 8 Pull the rope until the neck of the antenna is outside the mounting hole Step 9 Hand tighten the nut ...

Page 8: ...June 2017 8 Quick Start Guide Step 10 Remove the installation adapter Step 11 Tighten the mounting plate bolts 3x H2 mm ...

Page 9: ...k Start Guide 9 June 2017 3 0 Rotate the antenna toward vessel center Step 1 Slightly loosen the nut Step 2 Rotate the antenna The notch on top of the thread must be directed toward the vessel center 60 mm ...

Page 10: ...2017 10 Quick Start Guide Step 3 Tighten the nut Proceed with installation of scanner head See Rosemount 5708 Quick Start Guide and Rosemount 5708 Reference Manual for further installation information 60 mm ...

Page 11: ...Quick Start Guide 11 June 2017 ...

Page 12: ...Terms and Conditions of Sale can be found on the Terms and Conditions of Sale page The Emerson logo is a trademark and service mark of Emerson Electric Co Rosemount and Rosemount logotype are trademarks of Emerson All other marks are the property of their respective owners 2017 Emerson All rights reserved Europe Regional Office Emerson Automation Solutions Neuhofstrasse 19a P O Box 1046 CH 6340 Ba...