6.2
Viewing measurement data
6.2.1
Using the display panel
The Rosemount
™
5300 Series Radar Transmitter uses an optional display panel for
presentation of measurement data. When the transmitter is switched on, the display panel
presents information such as transmitter model, measurement frequency, software
version, communication type (HART
®
, FF), serial number, HART identification tag, setting of
write protection switch, and Analog Output settings.
When the transmitter is up and running, the display panel presents level, signal amplitude,
volume, and other measurement data depending on the display panel configuration (see
“Specifying display panel variables” on page 140
). The available LCD parameters are listed in
The display has two rows, the upper row shows the measurement value and the lower row
shows the parameter name and measurement unit. It toggles between the different
measurement values every two seconds. The lower row toggles between parameter name
and measurement unit each second.
Variables to be presented are configurable by using a Field Communicator, AMS
™
Device
Manager, DeltaV
™
or the Rosemount Radar Master software.
Figure 6-1. The Rosemount 5300 Series Display Panel
Error messages are listed in section
“LCD error messages” on page 193
6.2.2
Specifying display panel variables
It is possible to specify the variables to be presented on the display panel (LCD).
Using a Field Communicator
For a Field Communicator, the LCD settings are available with:
HART command [2, 2, 3].
F
OUNDATION
™
Fieldbus parameters:
TRANSDUCER 1100 > LCD_PARAMETERS.
Using Rosemount Radar Master (RRM)
The LCD tab in the
General
window lets you specify which variables to view on the Display
Panel screen:
3
Measurement value
Toggling between
measurement
parameter and
measurement unit