Operation and Maintenance
4.4.1
Flow/energy calculation verification (test calculation)
(Fully compensated mass and energy flow feature board only):
The Flow and Energy Calculation Verification Test allows the user to verify the flow configuration of the
Rosemount 3051SMV by entering expected values for the differential pressure, static pressure, and
process temperature variables. Under the
Flow/Energy Calculation Verification
heading, perform the
following steps:
1. Select the
Enable Test Calculation
button.
2. Select
Simulate DP
option. Select
Next
.
3. Select
DP Units
from the drop-down menu. Select
Next
.
4. Enter the DP Value corresponding to the desired flow rate simulation. Select
Next
.
for static pressure (Simulate AP/GP) and process temperature (Simulate PT), if
applicable.
6. Select
View Results
. Select
Next
. The simulated flow rate and corresponding flow properties will be
shown. Select
Next
.
7. Select
Exit
. Select
Next
. Leaving the
Enable Test Calculation
window automatically returns all process
variables fixed by the test calculation method to live process variable measurements.
4.4.2
Configuring fixed process variables
Under the
Fixed Process Variables
heading, the user may temporarily set the differential pressure, static
pressure or process temperature to a user defined fixed value for testing purposes. Once the user leaves
the
Configure Fixed Variable
method, the fixed process variable will be automatically returned to a live
process variable measurement.
4.4.3
Analog output loop test
Under the
Analog Output Verify
heading, a Loop Test can be performed by selecting the
Loop Test
button.
The loop test command verifies the output of the transmitter, the integrity of the loop, and the
operations of any recorders or similar devices installed in the loop.
Mass and energy flow Fast Keys
1, 2, 3
Mass and energy flow Fast Keys
1, 2, 4
Direct process variable output Fast Keys
1, 2, 3
Mass and energy flow Fast Keys
1, 2, 2
Direct process variable output Fast Keys
1, 2, 2