MODEL 2081 pH
SECTION 9.0
DESCRIPTION AND SPECIFICATIONS
9.3 PERFORMANCE SPECIFICATIONS
(Transmitter only)
Range:
0 to 14 pH
Output Scale:
Zero Suppression: Up to 13 pH units
Span: From 1 to 14 pH units
Accuracy:
± 0.02 pH, 0.1°C
Resolution:
0.01 pH, 0.1°C
Repeatability:
± 0.01 pH
Stability:
1%/ year at 25°C
Temperature Effect:
0.02% F.S./°C
Vibration Effect:
± 1.0% of F.S. per SAMA PMC
31.1, Section 5.3, Condition 2.
EMI/RFI:
EN50081-1
EN50082-2
9.4 PHYSICAL SPECIFICATIONS
Electrical Connections:
1/2 in. NPT.
Model 275 SMART FAMILY Interface
connections permanently fixed to terminal block.
The terminal block also has removable
connectors for bare wire sensor connections.
Housing:
Epoxy-polyester painted over low-copper
aluminum. BUNA-N O-rings on cover.
Recommended Cable:
Transmitter to power supply
two-wire, 18 AWG, shielded, Belden 8760 or
equal (Rosemount Analytical PN 9200001).
Weight/Shipping Weight:
2.18 kg/2.68 kg (4.8 lb/5.9 lb)
9.2 FUNCTIONAL SPECIFICATIONS
Inputs:
mV (for pH); resistance (for temperature)
Output:
Two-wire 4-20 mA with superimposed HART
digital signal.
Power Supply and Load Requirements:
See
graph
below. A minimum loop resistance (load) of 250
ohms and minimum power supply voltage of 18
volts DC is required for digital communication.
Local Indication:
Four digit LCD
Ambient Temperature:
–30 to 70°C (–22 to 158°F)
Failure Mode Alarm:
The analog signal will be driven
either below 4 mA or above 20 mA (user-
selectable) to alert the user.
Transmitter Security:
Security activation (by internal
jumper) prevents changes to the transmitter
configuration from the local interface.
Relative Humidity :
0-95%
Automatic Temperature Compensation:
Pt100 RTD
–15 to 115°C (5 to 239°F)
Enclosure:
NEMA 4X (IP65) , weatherproof and
corrosion-resistant NEMA 7B, explosion proof
Hazardous Area Classification:
Explosion Proof:
FM:
Class I, Div. 1, Groups B, C & D
Class II, Div. 1, Groups E, F, & G
Class III, Div. 1
CSA:
Class I, Div. 1, Groups C& D
Class I, Div. 2, Groups A, B, C & D
Class II, Div. 2, Groups E, F & G
Class III, Div. 1
Intrinsic Safety:
FM:
Class I, II & III, Div. 1
T4 T AMB= 40°C; T3AT AMB= 70°C
CSA:
Class I, Div. 1
T 3C T AMB=40°C; T3 T AMB=80°C
CENELEC: EEx ia IIC
T5 Tamb=40°C T4 Tamb=80°C
Non-Incendive:
FM:
Class I, Div. 2, Groups A, B, C & D
CSA:
Class I, Div. 2, Groups A, B, C & D T5
(Tamb=40°C)
LOAD/POWER SUPPLY REQUIRE-
MENTS
12 VDC
18 VDC
42.4 VDC @ ZERO LOAD
LIFT OFF
42.4 VDC MAXIMUM
POWER SUPPLY VOLTAGE
1848 –
1800 –
1500 –
1000 –
500 –
250 –
0
–
Load (ohms)
WITHOUT HART COMMUNICATION
OPERATING
REGION