TF 200.02.0013_eng
16
Pre-approved cleaning agents
Tradename
Solution
Acticlor plus
1000ppm
S90 sanirens
1-3% peracid
FUTUR, Alkalisk rens
1% Solution [pH 9,5]
Disinfection
Caution!
Disinfect the product between uses to avoid cross contamination
Caution!
Do not disinfect the product when connected to mains and/or are in use with a patient
6.
Maintenance
Notification!
Failing to comply with periodic maintenance may result of degrading of product function and
safety.
Notification!
Ropox will make parts list and circuit diagrams available for maintenance and repair beyond
what is described in this manual. Repair beyond what is described in this manual may only be
done by Ropox designated service personnel.
6.1 Periodic maintenance of bed guards
If the product is used in a humid or dry atmosphere the locking pin may be prone to failure from lubricants
drying. Besides checking the function of the locking mechanism before every use, a periodic lubrication of
the locking pin every 6 month is recommended.
6.2 Periodic charging of battery
The battery must be periodically charged minimum every 6 months to sustain capacity. If the product is left
unused or the battery is disconnected this may result degrading of the battery capacity. Charging the battery
minimum every 6 months will mitigate this.