TF 200.02.0013_eng
24
Phenomenon
Test method
Immunity test level
Immunity to conducted
disturbances induced
by RF fields
– AC
power ports
EN 61000-4-6
3 V (6 V in ISM bands and amateur radio bands
a
)
0.15-80 MHz
80 % AM 1 kHz
Immunity to conducted
disturbances induced
by RF fields
– I/O
SIO/SOP ports
EN 61000-4-6
3 V (6 V in ISM bands and amateur radio bandsa)
0.15-80 MHz
80 % AM 1 kHz
Power frequency
magnetic field immunity
EN 61000-4-8
30 A/m
50 Hz and 60 Hz
Test as tabletop equipment, for at least 1 minute in each
direction
Voltage dips, short
interruptions and
voltage variations
immunity
EN 61000-4-11
0 % U
T
; 0.5 cycle at 0°, 45°, 90°, 135°, 180°, 225°, 270° and
315°
0 % U
T
; 1 cycle at 0°
70 % U
T
; 10 cycles at 0°
0 % U
T
; 250 cycles at 0°
a) The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795 MHz; 13.553 MHz to
13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio bands between 0.15 MHz and 80 MHz
are 1.8 MHz to 2.0 MHz. 3.5 MHz to 4.0 MHz. 5.3 MHz to 5.4 MHz. 7 MHz to 7.3 MHz. 10.1 MHz to 10.15 MHz. 14 MHz to 14.2 MHz.
18.07 MHz to 18.17 MHz. 21.0 MHz to 21.4 MHz. 24.89 MHz to 24.99 MHz. 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.