Page 14
© ROPOX 2019
10. Cleaning/maintenance
10.1 Cleaning of frame
Clean the frame with a wrung cloth with lukewarm water and a mild detergent. Do not use
enchants or abrasives or grinding cloths, brushes or sponges
Dry the frame after cleaning.
10.2 Maintenance
The frame is maintenance-free and the moving parts have been lubricated for life. For reasons
of safety and reliability we recommend inspection of the frame once a year:
➢
Check that all bolts have been tightened.
➢
Check that the table moves freely and without problems from bottom to top position.
➢
Resetting of frame:
Press the DOWN button and adjust the frame to bottom position.
Press the DOWN button again and keep it pressed for 5 sec. to reset the control unit.
Check that the movement is smooth and even. Make sure that the mains cable moves
freely.
➢
Check that water connection hoses and draining hoses, if available, are tight and
functional.
➢
Check that cable to hotplate, if available, is undamaged and may move freely.
➢
Check that all cables have been mounted correctly and are undamaged.
After each inspection, fill in the service schedule. See 10.3
Only use Ropox original spare parts for replacement of parts. If other spare parts are used, the
right to complain may become void.
WARNING
The frame must NOT be connected to mains voltage during cleaning. Do
NOT flush electrical components with water.
Inspections, service and repairs must be carried out by competent personnel.