Electromagnetic compatibility
10.1
Suitable Environments
The device is suitable for use at home, at daycare centres, at day centres for persons with
physical or mental disabilities or at hospitals except near active HF surgical equipment and
the RF shielded room of an medical electrical system for magnetic resonance imaging,
where the intensity of electromagnetic disturbances is high.
10.2
Adjacent and stacked use
Warning !
Use of this equipment adjacent to or stacked with other equipment should be
voided because it could result in improper operation. If such use is necessary,
this equipment and the other equipment should be observed to verify that they
are operating normally
10.3
Cables
Cable component
Length
Power cable
1,2m
Hand control spiral cable
Wireless
Warning !
Use of accessories, transducers and cables other than those specified or
provided by the manufacturer of this could result in increased electromagnetic
emissions or decreased electromagnetic immunity of this equipment and result in
improper operation.
10.4
RF portable equipment
Warning !
Portable RF communications equipment (including peripherals such as antenna
cables and external antennas) should be used no closer than 30 cm (12 inches)
to any part of the device, including cables specified by the manufacturer.
Otherwise, degradation of the performance of this equipment could result.
Summary of Contents for 40-45027
Page 8: ...TF 200 06 0006_ENG 8 3 8 Product dimensions ...
Page 13: ...TF 200 06 0006_ENG 13 Mount the bellow as shown below for correct flow ...
Page 17: ...TF 200 06 0006_ENG 17 g Mount outer cover 2x ...
Page 18: ...TF 200 06 0006_ENG 18 h Shorten Ferrule and flush pipe to length and mount them ...
Page 23: ...TF 200 06 0006_ENG 23 ...
Page 24: ...TF 200 06 0006_ENG 24 d Slide in the cover and end rubber cap ...
Page 25: ...TF 200 06 0006_ENG 25 ...