![Ronde & Schwarz 1424.6721K02 User Manual Download Page 78](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980078.webp)
Remote control commands
R&S
®
NRP18S-xx
78
User Manual 1178.3686.02 ─ 07
[SENSe<Sensor>:]TIMing:EXCLude:STARt
<exclude_start>
Available in burst average and timeslot modes.
Sets a time interval at the beginning of bursts that is excluded from the measurement.
See
.
Parameters:
<exclude_start>
Range:
0.0 to 1.0
*RST:
0.0
Default unit: Seconds
[SENSe<Sensor>:]TIMing:EXCLude:STOP
<exclude_stop>
Available in burst average and timeslot modes.
Sets a time interval at the end of bursts that is excluded from the measurement. See
.
Parameters:
<exclude_stop>
Range:
0.0 to 1.0
*RST:
0.0
Default unit: Seconds
8.8.4
Configuring corrections
It is possible to set some parameters that compensate for a change of the measured
signal due to fixed external influences.
.............................................................................................78
........................................................................................... 80
..............................................................................................82
Using the S-Parameters program
........................................................................... 84
8.8.4.1
Duty cycle corrections
The duty cycle is the percentage of one period during which the signal is active, when
pulse-modulated signals are corrected. The duty cycle is only evaluated in the continu-
ous average mode.
Remote commands:
[SENSe<Sensor>:]CORRection:DCYCle
............................................................................ 78
[SENSe<Sensor>:]CORRection:DCYCle:STATe
.................................................................. 79
[SENSe<Sensor>:]CORRection:DCYCle
<duty_cycle>
Available in continuous average mode.
Configuring basic measurement parameters