Remote control commands
R&S
®
NRP18S-xx
64
User Manual 1178.3686.02 ─ 07
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:SIZE
<count>
Sets the size of the result buffer in continuous average mode.
You can enable the buffer using
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:
.
Parameters:
<count>
Range:
1 to 8192
*RST:
1
Example:
BUFF:SIZE 1
See
Chapter 9.3, "Performing a buffered continuous average
[SENSe<Sensor>:][POWer:][AVG:]BUFFer:STATe
<state>
Enables or disables the buffered continuous average mode. If the buffer mode is
enabled, all results generated by trigger events are collected in the sensor until the buf-
fer is filled.
You can set the size of the buffer using
.
Parameters:
<state>
ON | OFF
*RST:
OFF
Example:
BUFF:STAT OFF
[SENSe<Sensor>:][POWer:][AVG:]FAST
<state>
Enables or disables a fast unchopped continuous average measurement. If enabled,
the average count is enforced to 1, and any setting for average count is silently
ignored.
You can increase the measurement accuracy by increasing the duration of the sam-
pling window using:
[SENSe<Sensor>:][POWer:][AVG:]APERture
.
The fast measurement setting delivers up to 100
000
measurements per second with-
out any blind time over randomly long time periods.
See also
"Calculating the measurement time"
Parameters:
<state>
*RST:
OFF
Example:
FAST ON
See
Chapter 9.2, "Performing the fastest measurement in con-
Configuring the measurement modes