![Ronde & Schwarz 1424.6721K02 User Manual Download Page 107](http://html1.mh-extra.com/html/ronde-and-schwarz/1424-6721k02/1424-6721k02_user-manual_1484980107.webp)
Remote control commands
R&S
®
NRP18S-xx
107
User Manual 1178.3686.02 ─ 07
●
Chapter 10.2, "Status reporting system"
Contents:
General status register commands
....................................................................... 107
................................................................................. 107
........................................................................................108
.................................................................................. 108
Controlling the negative transition part
..................................................................108
Controlling the positive transition part
...................................................................109
8.13.1
General status register commands
................................................................................................ 107
STATus:PRESet
Resets the edge detectors and
ENABle
parts of all registers to a defined value.
Usage:
Event
STATus:QUEue[:NEXT]?
Queries the most recent error queue entry and deletes it.
Positive error numbers indicate sensor specific errors. Negative error numbers are
error messages defined by SCPI.
If the error queue is empty, the error number 0, "No error", is returned.
Usage:
Query only
8.13.2
Reading the CONDition part
STATus:DEVice:CONDition?
STATus:OPERation:CALibrating:CONDition?
STATus:OPERation:CONDition?
STATus:OPERation:LLFail:CONDition?
STATus:OPERation:MEASuring:CONDition?
STATus:OPERation:SENSe:CONDition?
STATus:OPERation:TRIGger:CONDition?
STATus:OPERation:ULFail:CONDition?
STATus:QUEStionable:CALibration:CONDition?
STATus:QUEStionable:CONDition?
STATus:QUEStionable:POWer:CONDition?
STATus:QUEStionable:WINDow:CONDition?
Usage:
Query only
Using the status register