Remote Commands
R&S
®
FPL1000
953
User Manual 1178.3370.02 ─ 11
Example:
CAL:RES?
would return, e.g.
Total Calibration Status:
PASSED, Date (dd/mm/yyyy): 12/07/2004,
Time: 16:24:54,Runtime: 00.06
Usage:
Query only
Manual operation:
See
DIAGnostic:SERVice:INPut:PULSed:CFRequency
<Frequency>
This command defines the frequency of the calibration signal.
Before you can use the command, you have to feed in a calibration signal with
DIAGnostic:SERVice:INPut[:SELect]
Manual operation:
See
DIAGnostic:SERVice:INPut:PULSed:MCFRequency
<Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(
DIAGnostic:SERVice:INPut[:SELect]
Parameters:
<Frequency>
*RST:
7 GHz
Default unit: Hz
Example:
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
DIAGnostic:SERVice:INPut:RF[:SPECtrum]
<Bandwidth>
This command selects the bandwidth of the calibration signal.
Parameters:
<Bandwidth>
NARRowband | BROadband
NARRowband
Narrowband signal for power calibration of the frontend.
BROadband
Broadband signal for calibration of the IF filter.
DIAGnostic:SERVice:INPut[:SELect]
<Signal>
This command activates or deactivates the use of an internal calibration signal as input
for the R&S
FPL1000.
Parameters:
<Signal>
CALibration
Uses the calibration signal as RF input.
Configuring the R&S
FPL1000