Remote Commands
R&S
®
FPL1000
951
User Manual 1178.3370.02 ─ 11
HCOP:PAGE:MARG1:BOTT 20
HCOP:PAGE:MARG1:LEFT 20
HCOP:PAGE:MARG1:RIGH 20
HCOP:PAGE:MARG1:TOP 20
//Configure the use of optimized colors for printout
HCOP:CMAP:DEF2
//Set format of printout to PDF.
HCOP:DEV:LANG1 PDF
//Define file name of printout
MMEM:NAME 'C:\R_S\instr\user\MeasurementTestReport.pdf'
//Store pdf of printout to file
HCOP:IMM
10.10
Configuring the R&S
FPL1000
The remote commands required to set up the R&S
FPL1000 are described here.
Configuring the Reference Frequency
.................................................................. 951
.........................................................................................952
.....................................................................................955
............................................................................ 959
Remote Commands for Language Settings
..........................................................964
Configuring the Network and Remote Control
...................................................... 964
Checking the System Configuration
......................................................................968
....................................................................................... 972
10.10.1
Configuring the Reference Frequency
........................................................................................ 951
[SENSe:]ROSCillator:SOURce
<Source>
This command selects the reference oscillator.
If you want to select the external reference, it must be connected to the R&S
FPL1000.
Parameters:
<Source>
INTernal
The internal reference is used (10 MHz).
EXTernal
The external reference from the "REF INPUT 10
MHZ" connec-
tor is used; if none is available, an error flag is displayed in the
status bar.
Example:
ROSC:SOUR EXT
Configuring the R&S