Remote Commands
R&S
®
FPL1000
753
User Manual 1178.3370.02 ─ 11
[SENSe:]FREQuency:SPAN:FULL
This command restores the full span.
Manual operation:
See
[SENSe:]FREQuency:STARt
<Frequency>
This command defines a start frequency for measurements in the frequency domain.
Parameters:
<Frequency>
0 to (fmax - min span)
*RST:
0
Default unit: HZ
Example:
FREQ:STAR 20MHz
Manual operation:
See
See
[SENSe:]FREQuency:STOP
<Frequency>
This command defines a stop frequency for measurements in the frequency domain.
Parameters:
<Frequency>
min span to fmax
*RST:
fmax
Default unit: HZ
Example:
FREQ:STOP 2000 MHz
Manual operation:
See
See
10.8.1.2
Adjusting Settings Automatically
The commands required to adjust settings automatically in a remote environment are
described here.
[SENSe:]ADJust:CONFigure:LEVel:DURation
...................................................................754
[SENSe:]ADJust:CONFigure:LEVel:DURation:MODE
........................................................ 754
[SENSe:]ADJust:CONFigure:HYSTeresis:LOWer
.............................................................. 755
[SENSe:]ADJust:CONFigure:HYSTeresis:UPPer
............................................................... 755
[SENSe:]ADJust:CONFigure:TRIGger
..............................................................................755
..........................................................................................756
[SENSe:]ADJust:ALL
This command initiates a measurement to determine and set the ideal settings for the
current task automatically (only once for the current measurement).
Setting Basic Spectrum RF Measurement Parameters