Remote Commands
R&S
®
FPL1000
908
User Manual 1178.3370.02 ─ 09
CALCulate<n>:LIMit<li>:UNIT
<Unit>
This command defines the unit of a limit line.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Unit>
DBM | DBPW | WATT | DBUV | DBMV | VOLT | DBUA |
AMPere | DB | DBUV_M | DBUA_M | (unitless)
If you select dB as the limit line unit, the command automatically
turns the limit line into a relative limit line.
*RST:
DBM
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer[:DATA]
<LimitLinePoints>
This command defines the vertical definition points of an upper limit line.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FPL1000 either
adds missing values or ignores surplus values.
The unit depends on
on page 908.
*RST:
Limit line state is OFF
Default unit: DBM
Manual operation:
See
CALCulate<n>:LIMit<li>:UPPer:MARGin
<Margin>
This command defines an area around an upper limit line where limit check violations
are still tolerated.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Margin>
numeric value
*RST:
0
Default unit: dB