Remote Commands
R&S
®
FPL1000
905
User Manual 1178.3370.02 ─ 09
Parameters:
<LimitLinePoints>
Variable number of level values.
Note that the number of vertical values has to be the same as
the number of horizontal values set with
FPL1000 either
adds missing values or ignores surplus values.
The unit depends on
on page 908.
*RST:
Limit line state is OFF
Default unit: DBM
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:MARGin
<Margin>
This command defines an area around a lower limit line where limit check violations
are still tolerated.
Suffix:
<n>
.
irrelevant
<li>
Parameters:
<Margin>
numeric value
*RST:
0
Default unit: dB
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:MODE
<Mode>
This command selects the vertical limit line scaling.
Suffix:
<n>
.
<li>
Parameters:
<Mode>
ABSolute
Limit line is defined by absolute physical values.
The unit is variable.
RELative
Limit line is defined by relative values related to the reference
level (dB).
*RST:
ABSolute
Manual operation:
See
CALCulate<n>:LIMit<li>:LOWer:OFFSet
<Offset>
This command defines an offset for a complete lower limit line.
Setting Basic Spectrum RF Measurement Parameters