Remote Commands
R&S
®
FPL1000
811
User Manual 1178.3370.02 ─ 09
Manual operation:
See
See
SYSTem:SPEaker:MUTE
Temporarily disables the audio output via the built-in loudspeakers.
Example:
SYST:SPE:MUTE
Manual operation:
See
SYSTem:SPEaker:VOLume
<Volume>
This command defines the volume of the built-in loudspeaker for demodulated signals.
This setting is maintained for all applications.
The command is available in the time domain in Spectrum mode and in Analog Modu-
lation Analysis mode.
Note that you must switch the loudspeaker on first, using the
command.
Parameters:
<Volume>
Percentage of the maximum possible volume.
Range:
0 to 1
*RST:
0.5
Example:
SYST:SPE:VOL 0
Switches the loudspeaker to mute.
Manual operation:
See
See
11.8.6
Zooming into the Display
11.8.6.1
Using the Single Zoom
DISPlay[:WINDow<n>][:SUBWindow<w>]:ZOOM:AREA
.................................................... 811
DISPlay[:WINDow<n>][:SUBWindow<w>]:ZOOM[:STATe]
..................................................813
DISPlay[:WINDow<n>][:SUBWindow<w>]:ZOOM:AREA
<x1>,<y1>,<x2>,<y2>
This command defines the zoom area.
To define a zoom area, you first have to turn the zoom on.
Setting Basic Spectrum RF Measurement Parameters