The Spectrum Application (RF Measurements)
R&S
®
FPL1000
269
User Manual 1178.3370.02 ─ 09
For more information see
"Controlling V-Networks (LISN)"
LISN Type
Selects the network type and activates output to the network via the user port of the
R&S
FPL1000. The network type determines the supported phases (see
"Off" disables LISN control and output.
Remote command:
Phase
Selects the phase to be measured. Phase N and L1 are included in all four LISN.
Phase L2 and L3 are only included in four-line networks.
You can select one phase only for each measurement.
Remote command:
150 kHz Highpass
Enables or disables the use of an additional 150
kHz highpass filter to protect the
R&S
FPL1000 LISN from excessive input.
The filter is available for the ENV 216 network only.
Remote command:
INPut<ip>:LISN:FILTer:HPASs[:STATe]
Measurements and Results