The Spectrum Application (RF Measurements)
R&S
®
FPL1000
191
User Manual 1178.3370.02 ─ 09
1. Configure and perform an SEM measurement as described in
"How to Perform a Spectrum Emission Mask Measurement"
2. In the "Overview", select the "Analysis" button.
3. If necessary, change the "Decimal Separator" to "COMMA" for evaluation in other
languages.
4. Select the "Save" button.
5. In the file selection dialog box, select a storage location and filename for the result
file.
6. Select the "Save" button.
The file with the specified name and the extension
.dat
is stored in the defined
storage location.
7.2.7.7
Measurement Example: Multi-SEM Measurement
The following measurement example demonstrates an SEM measurement for a signal
with multiple sub blocks.
A programming example demonstrating a SEM measurement in a remote environment
is provided in
Chapter 11.6.7.10, "Example: SEM Measurement"
Test setup:
Signal generator settings (e.g. R&S
FPL1000 SMW):
Device
Standard
Center frequency
Level
Test model
SigGen 1
3GPP/FDD
900 MHz
0 dBm
1-16
SigGen 2
EUTRA/LTE
906.5 MHz
0 dBm
1_1_5MHz
Measurements and Results