The Spectrum Application (RF Measurements)
R&S
®
FPL1000
183
User Manual 1178.3370.02 ─ 09
Remote command:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>[:EXCLusive]
on page 681
To define all limits in one step:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:LIMit[:STATe]
PMin/ PMax
Defines the power limits for each power class. The first range always starts at -200
dBm (-INF) and the last range always stops at 200 dBm (+INF). These fields cannot be
modified. If more than one power class is defined, the value of "PMin" must be equal to
the value of "PMax" of the previous power class and vice versa.
Note that the power level can be equal to the lower limit(s), but must be lower than the
upper limit(s):
P
min
≦
P
<
P
max
Otherwise the ranges are corrected automatically.
Remote command:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MINimum
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:MAXimum
Sweep List
Switches to the "Sweep List" tab of the "Spectrum Emission Mask" dialog box and
focuses the "Limit Check" setting for the corresponding power class (1-4) in the refer-
ence range (see
Adding or Removing a Power Class
Adds a new power class at the end of the list or removes the last power class. After
adding or removing, the last power class is adapted to end at "+INF". Note that a maxi-
mum of four power classes are available.
Remote command:
CALCulate<n>:LIMit<li>:ESPectrum<sb>:PCLass<pc>:COUNt
Standard Files
Access
: "Overview" > "Select Measurement" > "Spectrum Emission Mask" > "Stan-
dard Files"
You can save the current measurement settings as a user-defined standard (XML file),
or load stored measurement settings. Furthermore, you can delete an existing settings
file.
Measurements and Results