Rev. 05
-
Page 6 of 12
-
09/10
5.3.3
Put the lip clip on the patient’s lip.
5.3.4
Insert the endodontic file into the root canal and connect the file clip to the
endodontic file.
Make sure that irrigation solution, blood or saliva are not overflowing
from the tooth cavity, dry the cavity before performing measurements.
5.3.5
RomiApex™ A-15
detects automatically beginning of measurements,
checks the quality of the electrical contact and the conductivity of the root
canal.
5.3.6
When the electrical contact is good, an audio signal of two beeps indicates
beginning of normal measurement cycle. Tooth image will stop blinking.
If there are no beeps, clean the file clip and the file, irrigate the canal, if
required, and continue the measurements.
It is not recommended to
continue measurements if the tooth image blinks.
5.3.7 Movement of the file inside the canal is reflected by the
DOWN
(Pic. 4) and
UP
(Pic. 5) arrows on the screen.
Pic. 4
Pic. 5