55
1. The left* motor has a short
circuit re: a battery
connection
Contact an authorised repairs
service.
1. Poor connection to the
right* motor
Check the connection to the motor
1. The right* motor has a
short circuit re: a battery
connection
Contact an authorised repairs
service.
1. The wheelchair has been
prevented from running by an
external signal, e.g. special
solution with contact.
The reason will depend on the
special solution model - Contact
ROLTEC or the supplier.
1. Joystick not centred
Turn the wheelchair off, centre the
joystick and turn the wheelchair
on again.
2. Joystick error.
Contact an authorised repairs
service.
1. System error.
Check all connections
+ control box beeping
1. Brakes are mechanically
disconnected
Reconnect the brakes,
see
section Mechanical
disconnection of brake
2. Poor connection to brake Check that the motor/brake
wiring is attached firmly in the
electronics box on the
wheelchair.
1. An excess of voltage
has occurred in the
control system.
This is normally caused by a
poor battery connection.
Check the connections to the
batteries.
1. Poor connection to
connector.
Check the cable connections
between the control box, seat
electronics and
bottom electronics.
2. Cable broken
Replace cable.
1. Electrical function fault.
Check whether one of the electrical
functions is working, check the
connections from the electronics of
the actuator that
does not work.
* If the programme has been changed, it could be the opposite motor.