ROHO Sensor Ready Cushion Operation Manual
9
Cleaning and Disinfecting
Cleaning and disinfecting are separate processes. Cleaning must precede disinfection.
Before use by a different individual: clean, disinfect, and check product for proper
functioning.
Cover Care:
Remove hook fastener pieces from bottom of cover prior to cleaning and
disinfecting. Wash with like colors.
To Clean the Cover:
Remove the cover. Machine wash in cold water (30°C)
with mild detergent, gentle cycle, or wipe clean with neutral detergent and
warm water. Tumble dry low.
To Disinfect the Cover:
Hand wash in cold water
(30°C) using 1 part household liquid bleach per
9 parts water. Keep the product wet with bleach
solution for 10 minutes, and then rinse thoroughly.
Or machine wash in hot water (60°C). Tumble dry low.
Follow safety guidelines on bleach container.
Cautions:
Valve(s) must be closed.
DO NOT allow water or cleaning solution to enter the cushion.
DO NOT machine wash or sanitize the cushion or machine dry the cushion.
Use of the following may cause damage to the cushion:
• abrasives (steel wool, scouring pads)
• caustic, automatic dishwasher detergents
• cleaning products that contain petroleum or organic solvents, including acetone,
toluene, Methyl Ethyl Ketone (MEK), naphtha, dry cleaning fluid, adhesive remover
• oil-based lotions, lanolin, or phenolic-based disinfectants
• ozone gas; ultraviolet light cleaning methods
To Clean the Cushion:
Remove the cover and close the inflation valve. Hand wash, using
mild liquid hand soap, hand dishwashing liquid, laundry detergent, or a multipurpose
detergent (Follow product label instructions.) Gently scrub all cushion surfaces, using
a soft plastic bristle brush, a sponge or a cloth. Rinse with clean water. Air dry. DO NOT
expose to direct sunlight.
To Disinfect the Cushion:
Repeat cleaning instructions, using 1 part household liquid
bleach per 9 parts water. Keep the product wet with the bleach solution for 10 minutes.
Rinse thoroughly with clean water. Air dry.
Follow safety guidelines on bleach container.
Note: Most germicidal disinfectants are safe if used following the disinfectant
manufacturer’s dilution directions.
30C
30C
1:9
+
or