Rohde & Schwarz ZVH4 Manual Download Page 6

Instrument Declassification

R&S

®

ZVH

6

Instrument Security Procedures 1178.4324.02 ─ 02

Factory default settings (display type and contrast, installed options, baud rate, lan-
guage, length unit etc.)

User general settings (display type and contrast, installed options, baud rate, lan-
guage, length unit etc.)

User data (cable models, transducers, limit lines, channel tables, user standards)

User calibration data (DTF calibration data, tracking generator calibration data,
etc.)

Saved user data sets

The flash memory does not lose its memory when power is removed. Therefore, the
memory device must be cleared before leaving the secured area.

Sanitization procedure:

 Master preset procedure (see 

Chapter 5, "Instrument Declas-

sification"

on page 6)

The instrument firmware and factory default settings section of the flash memory are
not erased by the master preset procedure. The instrument firmware and factory
default settings memory section does not hold any user data.

5

 

Instrument Declassification

Instrument declassification requires firmware 1.70 or higher.

Before you can remove the R&S

 

ZVH Cable and Antenna Analyzer from a secured

area (for example to perform service or calibration), all classified user data needs to be
removed. You can declassify the R&S

 

ZVH as follows:

1. Turn off the R&S

 

ZVH. This will sanitize the volatile memory.

2. To sanitize the internal Flash memory, perform the following steps:

a) Make sure, that no USB mass memory device is connected.
b) Press the front panel buttons [PRESET] and softkey [5] and hold them while

switching on the instrument again.

After a few seconds, the sanitizing procedure starts.

Sanitizing is indicated by the message "Secure Formatting Flash, please wait!" on
the instrument's screen. The sanitizing procedure takes approximately 8 minutes.

Afterwards, the instrument reboots. Since permanent adjustment values are not loca-
ted in instrument settings and user data section of the flash, the validity of the
R&S

 

ZVH Cable and Antenna Analyzer’s calibration is maintained throughout the sani-

tization.

Following these steps removes all user data from the R&S

 

ZVH Cable and Antenna

Analyzer. The instrument can now leave the secured area.

Summary of Contents for ZVH4

Page 1: ...R S ZVH Cable and Antenna Analyzer Instrument Security Procedures Instrument Security Procedures Version 02 1178432402 Ü H2 ...

Page 2: ...information leaves with the test equipment Security concerns can arise when devices need to leave a secured area e g to be calibrated or serviced This document describes the types of memory and their usage in the R S ZVH It pro vides a statement regarding the volatility of all memory types and specifies the steps required to declassify an instrument through memory clearing or sanitization proce du...

Page 3: ...ry sanitization procedures described in this document are designed for cus tomers who need to meet the requirements specified by the US Defense Security Ser vice DSS These requirements are specified in the Clearing and Sanitization Matrix in Section 14 1 16 of the ISFO Manual for the Certification and Accreditation of Classi fied Systems under the NISPOM Instrument declassification The term instru...

Page 4: ...on on page 6 Flash main board 4 Mbyte or 8 Mbyte Instrument firmware Factory default settings User general settings User data User calibration data Saved user data sets Non volatile Yes Sanitize internal memory procedure see Chapter 5 Instrument Declas sification on page 6 4 1 Volatile Memory The volatile memory in the instrument loses its contents as soon as power is removed from the instrument T...

Page 5: ... EEPROM contains information related to the instal led hardware such as board serial number product options and calibration correction data The EEPROM does not hold user data nor can the user access the EEPROM storage Sanitization procedure None required no user data SRAM The main board of the R S ZVH Cable and Antenna Analyzer has one SRAM memory device which holds 1 Mbyte or 8 Mbyte depending on...

Page 6: ... Instrument declassification requires firmware 1 70 or higher Before you can remove the R S ZVH Cable and Antenna Analyzer from a secured area for example to perform service or calibration all classified user data needs to be removed You can declassify the R S ZVH as follows 1 Turn off the R S ZVH This will sanitize the volatile memory 2 To sanitize the internal Flash memory perform the following ...

Page 7: ...djustment values required to maintain the validity of the R S ZVH s calibration Therefore performing the memory clearing procedures described above does not affect the validity of the instrument s calibration 2019 Rohde Schwarz GmbH Co KG Mühldorfstr 15 81671 München Germany Phone 49 89 41 29 0 Fax 49 89 41 29 12 164 Email info rohde schwarz com Internet www rohde schwarz com Subject to change Dat...

Reviews: