![Rohde & Schwarz ZNL Series User Manual Download Page 61](http://html.mh-extra.com/html/rohde-and-schwarz/znl-series/znl-series_user-manual_1477893061.webp)
Analysis
R&S
®
ZNL
61
User Manual 1178.5989.02 ─ 06
Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6
..............................................61
Predefined Trace Settings - Quick Config
.................................................................... 63
Trace 1 / Trace 2 / Trace 3 / Trace 4 (Softkeys)
............................................................63
Trace 1 / Trace 2 / Trace 3 / Trace 4 / Trace 5 / Trace 6
Selects the corresponding trace for configuration. The currently selected trace is high-
lighted.
Remote command:
Selected via numeric suffix of:
TRACe<1...6>
commands
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>[:STATe]
TraceMode
Defines the update mode for subsequent traces.
"Clear/ Write"
Overwrite mode (default): the trace is overwritten by each sweep.
"Max Hold"
The maximum value is determined over several sweeps and dis-
played. The R&S
ZNL saves each trace point in the trace memory
only if the new value is greater than the previous one.
"Min Hold"
The minimum value is determined from several measurements and
displayed. The R&S
ZNL saves each trace point in the trace memory
only if the new value is lower than the previous one.
"Average"
The average is formed over several sweeps.
The
determines the number of averaging pro-
cedures.
Trace Settings