Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
124
User Manual 1178.5989.02 ─ 06
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]
<Range>
This command defines the display range of the y-axis (for all traces).
Note that the command works only for a logarithmic scaling. You can select the scaling
with
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Range>
Range:
1 dB to 200 dB
*RST:
100 dB
Default unit: HZ
Example:
DISP:TRAC:Y 110dB
Manual operation:
See
See
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:AUTO ONCE
Automatic scaling of the y-axis is performed once, then switched off again (for all
traces).
Suffix:
<n>
.
<t>
irrelevant
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y[:SCALe]:MODE
<arg0>
This command selects the type of scaling of the y-axis (for all traces).
When the display update during remote control is off, this command has no immediate
effect.
Suffix:
<n>
.
<w>
subwindow
<t>
irrelevant
Parameters:
<arg0>
ABSolute
absolute scaling of the y-axis
RELative
relative scaling of the y-axis
*RST:
ABSolute
Example:
DISP:TRAC:Y:MODE REL
Manual operation:
See
Configuring I/Q Analyzer Measurements