Remote Commands to Perform Measurements with I/Q Data
R&S
®
ZNL
125
User Manual 1178.5989.02 ─ 06
DISPlay[:WINDow<n>]:TRACe<t>:Y[:SCALe]:RPOSition
<Position>
This command defines the vertical position of the reference level on the display grid
(for all traces).
The R&S
ZNL adjusts the scaling of the y-axis accordingly.
Suffix:
<n>
.
<t>
irrelevant
Parameters:
<Position>
0 PCT corresponds to the lower display border, 100% corre-
sponds to the upper display border.
*RST:
100 PCT = frequency display; 50 PCT = time dis-
play
Default unit: PCT
Example:
DISP:TRAC:Y:RPOS 50PCT
Manual operation:
See
DISPlay[:WINDow<n>][:SUBWindow<w>]:TRACe<t>:Y:SPACing
<ScalingType>
This command selects the scaling of the y-axis (for all traces, <t> is irrelevant).
Suffix:
<n>
.
<w>
subwindow
<t>
Parameters:
<ScalingType>
LOGarithmic
Logarithmic scaling.
LINear
Linear scaling in %.
LDB
Linear scaling in the specified unit.
PERCent
Linear scaling in %.
*RST:
LOGarithmic
Example:
DISP:TRAC:Y:SPAC LIN
Selects linear scaling in %.
Manual operation:
See
Configuring I/Q Analyzer Measurements