Command Reference
R&S
®
ZNB/ZNBT
952
User Manual 1173.9163.02 ─ 38
Example:
Suppose that a full two-port calibration for ports 1 and 2 and
channel 1 has been performed, and that a DUT with two bal-
anced ports is connected. The analyzer measures an arbitrary
mixed mode S-parameter.
MMEM:STOR:TRAC:PORT 1, 'Test_CIMP.s2p',
COMPlex, CIMPedance 1,2
Calculate all single-ended S-parameters, renormalize them to
the common target impedance and store them to a two-port
Touchstone file.
MMEM:STOR:TRAC:PORT 1, 'Test_PIMP.s2p',
COMPlex, PIMPedance 1,2
Calculate all single-ended S-parameters, renormalize them to to
the individual port reference impedances and store them to a
two-port Touchstone file.
Usage:
Setting only
Manual operation:
See
7.3.13 OUTPut Commands
The
OUTPut...
commands control the characteristics of the analyzer’s output ports.
......................................................................................... 953
OUTPut:UPORt:ECBits
<Boolean>
Defines the usage of pins pins 16 to 19 of the USER PORT connector.
Parameters:
<Boolean>
ON
–
channel bits 4 to 7
OFF
–
drive port 1 to 4
*RST:
ON
Example:
See
Manual operation:
See
OUTPut<Ch>[:STATe]
<Boolean>
Turns the internal source power at all ports and the power of all external generators on
or off.
Suffix:
<Ch>
.
Channel number. This suffix is ignored; the setting is valid for all
channels.
Parameters:
<Boolean>
ON | OFF - switch the power on or off.
*RST:
ON
SCPI Command Reference