Command Reference
R&S
®
ZNB/ZNBT
681
User Manual 1173.9163.02 ─ 38
Return values:
<Data>
Response values either in ASCII or block data format, depend-
ing on the current
Example:
See
Usage:
Query only
CALCulate:DATA:DALL?
<Format>
Reads the current response values of all data traces of the current recall set. Use
to query data traces and memory traces.
Query parameters:
<Format>
FDATa | SDATa | MDATa
Output format for the S-parameter data, see
Return values:
<Data>
Response values either in ASCII or block data format, depend-
ing on the current
Example:
Analogous to
Usage:
Query only
CALCulate:DATA:TRACe
<TraceName>, <Format>, <Data>...
The query gets the trace data of an arbitrary (not necessarily the active) trace, refer-
enced by its trace name <TraceName>.
The "set direction" allows to import formatted or unformatted trace data to an existing
trace (data access point 3 or 2 in
Note
●
Unformatted trace data (SDATa) can only be imported to memory traces
●
Formatted trace data (FDATa) can only be imported to "live" traces if the related
channel is in single sweep mode (
).
Before the import, the target trace must be prepared according to the settings used
during export. Any request for new data from the hardware ("Restart Sweep" in sin-
gle sweep mode or switching to continuous sweep mode) discards imported data
and switches back the display to measured data.
Parameters:
<TraceName>
String parameter containing the trace name
<Format>
FDATa | SDATa | MDATa | NCData | UCData
Data format; see
.
<Data>
Trace data for
FDATa | SDATa
import either in ASCII or block
data format, depending on the current
setting.
The column order must match the one used during export.
SCPI Command Reference