A PAE measurement involves the following steps:
1. Select the "Measurement Type", according to the properties of your DC power
source.
2. Enter the parameters for the selected measurement type.
3. Establish the test setup (including the RF connection of the DUT and the DC
INPUT connection(s)) as shown in the circuit diagram.
4. Close the "Power Added Efficiency (PAE)" dialog and observe the result in the
measurement diagram.
Measurement Type
Selects the PAE measurement type, depending on the DC power source in use (DC
source providing a constant, known voltage at (possibly) variable current / DC source
providing a constant, known current at (possibly) variable voltage / DC source with
variable (unknown) current and power). The three measurement types involve different
test setups, input parameters, and DC voltage measurements. For an overview see
Note:
The measurement types for "Constant Voltage Source" and "Voltage and Cur-
rent" require an additional non-zero, known precision resistor R.
Remote command:
Formula
Shows the formula for the calculation of the DC supply power P
DC
, depending on the
selected measurement type. The parameters R, U
0
, and/or I
0
must be entered in
accordance with the properties with the DC power source and precision resistor in use.
Remote command:
[SENSe<Ch>:]PAE:PARameters:I
[SENSe<Ch>:]PAE:PARameters:R
[SENSe<Ch>:]PAE:PARameters:U
Connector buttons
Define the VNA test ports for the RF input and output signal (In / Out) at the DUT and
the DC connector(s) used to determine the DC supply power. The "Voltage and Cur-
rent" measurement type requires two simultaneous DC voltage measurements. The
DC range of each DC INPUT connector in use should be selected according to the
(estimated) input voltage range; see
"DC 1 / DC 2 / DC 3 / DC 4 / Source Port"
Remote command:
[SENSe<Ch>:]PAE:DCINput:MAIN
[SENSe<Ch>:]PAE:DCINput:SECondary