GUI Reference
R&S
®
ZNB/ZNBT
403
User Manual 1173.9163.02 ─ 38
●
"Segment" means that each trigger event starts the next sweep segment within the
current sweep. If a sweep type other than
is active, this option is equiv-
alent to "Sweep".
Relation with other sweep settings
Some sweep settings are logically incompatible with a particular selection of the trig-
gered measurement sequence:
●
If a
sweep is performed, the sequence is always a sweep.
●
"Alternated"
only makes sense if the triggered measurement
sequence comprises more than one sweep point. If "Point" or "Partial Measure-
ment" is selected, "Alternated" mode is switched off and vice versa.
Note:
The period of the trigger events should be adjusted to the triggered measure-
ment sequence. If the analyzer receives a trigger event while the last sequence is still
running, the R&S
ZNB/ZNBT skips the trigger event and generates a message.
Remote command:
Delay
Specifies a delay time between the trigger event and the start of the measurement.
The "Delay" entered must be zero or positive, so that the trigger event precedes the
start of the measurement (post-trigger).
If "Multiple Triggers" is active the "Delay" can be selected individually using the
.
Remote command:
TRIGger<Ch>[:SEQuence]:HOLDoff
Slope/Level
Specifies the "External" trigger mode in detail.
●
Rising Slope / Falling Slope
means that the rising or falling slope of every exter-
nal trigger pulse can trigger a single measurement sequence.
●
High Level / Low Level
means that the analyzer performs a free run measurement
as long as the external trigger signal is high / low. The measurement is discontin-
ued when the trigger signal changes to low / high.
Remote command:
5.6.1.2
Trigger Manager Dialog
The "Trigger Manager" dialog defines individual trigger sources and delays for the trig-
gered measurement sequences. The dialog is available and its settings are valid if the
analyzer is configured for "Multiple Triggers" (see
"FreeRun / External / Manual / Multi-
Access:
CHANNEL – TRIGGER > "Trigger" > "Trigger Manager..."
Background information
Refer to
Chapter 4.1.4.1, "Partial Measurements and Driving Mode"
Trigger Softtool