GUI Reference
R&S
®
ZNB/ZNBT
303
User Manual 1173.9163.02 ─ 38
Calibrate all Channels
Check this box to apply the acquired correction data to all channels in the active recall
set. Leave it unchecked (preset setting) to apply them only to the active channel.
Note that this option is available only if the active recall set contains multiple channels.
Remote command:
[SENSe:]CORRection:COLLect:CHANnels:ALL
Add
Adds a new system error correction.
The "Add" button opens the
dialog without pre-selected ports.
Delete
Deletes the selected calibration from the list of configured calibrations.
Modify
Edits the selected system error correction: opens the
dialog
with the corresponding ports and calibration type pre-selected.
Next
.
Inactive as long as no SMARTerCal is defined.
Define Calibration Dialog
In MultiCal mode the "Define Calibration" dialog allows to define the individual SMAR-
TerCals by selecting the test ports to be calibrated as well as the calibration unit, char-
acterization, calibration type, power meter and power port to be used.
Access:
CHANNEL – CAL > "Start Cal" > "SMARTerCal" – "Start... (Cal Unit)" >
"Add..." / "Modify..."
●
The "Multiple Cal in Calibration Wizard" (MultiCal) feature is disabled by default
and has to be activated in the
tab of the System Config dialog.
●
For background information on calibration types see
Cal Softtool